PM150-CO2 Portable Gas Monitor
REV – 1.0
[34-2900-0211-5]
9
|
P a g e
5.3
Settings Mode
In the settings mode, users can adjust set points, perform calibration, and reset previous values.
•
To enter the setting mode, press and hold the Function key & power button simultaneously for three
seconds. The following menu ALr
→
CAL
→ C
Lr MAX
→ C
Lr and STEL, TWA
→
UnIt
→
InIt
→
tESt is
displayed.
•
To move to the next menu, press the Function key.
•
To enter the menu, press and hold down the power button.
5.4
Setting Mode Symbols
Setting
Submenu
LCD
Action
ALr
1
st
Alarm
2
nd
Alarm
1
st
Alarm concentration setting
2
nd
Alarm concentration setting
CAL
ZERO (N2) SPAN
(CO2)
Zero (N2) calibration Span (CO2)
calibration
CLr MAX
-
Delete maximum alarm Concentration
CLr STEL,
TWA
-
Delete STEL and TWA Concentration
Unit
%vol / ppm
Concentration unit conversion
Init
-
Reset
Test
Self
Bts
Self-test
Bump test