PM150-CO2 Portable Gas Monitor 

REV – 1.0 

[34-2900-0211-5] 

15

 

|

 

P a g e

 

 

 

8.3

 

Span Calibration 

 

1.

 

Assemble the calibration equipment with span gas cylinder. See Figure 8-1 

2.

 

In the “CAL” mode, press the Function key to move to the “SPAn” calibration. 

3.

 

Attach the calibration cap to the detector. 

4.

 

Press and hold the power button for three (3) seconds to start the span calibration. Once the countdown timer 

has started, open the valve on the regulator to start the flow of gas. 

5.

 

After 90 seconds, when the span calibration is successful, success message (

V

) appears. 

If span calibration fails, fail message (

X

) appears. 

Note

: Press the power button during the 90 second countdown to abort the span calibration. 

 

CAUTION

 

Do not change the calibration concentration without a safety managers authorization. 

Use the regulator with a flow rate of 0.2 LPM (Liters per minute). 

 

 

Содержание AimSafety PM150-CO2

Страница 1: ...AimSafety PM150 CO2 Carbon Dioxide Gas Monitor User Instructions ...

Страница 2: ...tion 10 6 2 Adjust alarm set points 10 6 3 Data Log 10 7 Menu Screens 11 7 1 Menu Flow Chart 11 7 2 Clear Max Peak Values 12 7 3 Clear STEL and TWA 12 7 4 Adjust Unit 12 7 5 Factory Reset 12 7 6 Self Test 13 8 Calibration 14 8 1 Calibration Gas Values 14 8 2 Zero N2 Calibration 14 8 3 Span Calibration 15 9 Software Manager 16 9 1 Software Overview 16 9 1 1 Read 17 9 1 2 Write 17 9 1 3 Calibration ...

Страница 3: ...irt and debris that could block the area where gas is to be detected Do not clean and rub the LCD screen of the products with a dry cloth or hands in hazardous environment to prevent static electricity buildup If the CO2 concentration reaches 0 ppm a calibration should be performed CAUTIONS Before operating this device please read and understand this manual This device is not an analyzer but a gas...

Страница 4: ...t Diffusion type Display LCD display Audible 90dB at 10cm Warning Lamp Red Flashing LEDs Vibration VibrationAlarm Battery Rechargeable Li ion polymer power supply unit 500mAh Charging Time 100 minutes from empty battery to fully charge based on power off state and charging with adaptor Temperature 20ºC to 50ºC Humidity 5 to 95 RH Non condensing Case Rubber Enclosure Accessories Calibration Cap Cha...

Страница 5: ...Calibration Setting Success Calibration Setting Failure Measurement Unit Standard Gas Calibration Remaining Battery Alarm Condition 1st Alarm Display 2nd Alarm Display Max Value Short Term Exposure Limit Time Weighted Average Test Success Test Fail Over Limit End of Test LCD display Alarm LED Buzzer Power Button Function Key Gas sensor ...

Страница 6: ...ears and thedetector performs the self diagnostic test for 10 seconds After self test is successful the detector countdown is displayed for 10 seconds Allow the detector to stabilize for 90 seconds until warm up message is no longerdisplayed The detector will then display the current CO2 concentration 4 2 Error Messages If stabilization has failed Error mark is displayed and measuring mode will no...

Страница 7: ...before performing a Bump Test The monitor must be calibrated if it fails a Bump Test To perform the bump test Prepare a gas cylinder with a CO2 concentration that exceeds the low and high alarm set points Press and hold the Function key and power button for three seconds in the measurementmode Press the Function key until TEST is displayed and press the power button two 2 seconds to enter the mode...

Страница 8: ...oint TWA alarm set point Firmware version Calibration concentration Calibration interval remaining date Bump test interval remaining date To move to the next menu press the Function key At the last step press Function key or do not push any button for 10 seconds the device will return to the Measuring Mode Max Value Measured STEL value Measured TWAvalue LowAlarm Value Setting HighAlarm Value Setti...

Страница 9: ...and STEL TWA UnIt InIt tESt is displayed To move to the next menu press the Function key To enter the menu press and hold down the power button 5 4 Setting Mode Symbols Setting Submenu LCD Action ALr 1st Alarm 2nd Alarm 1st Alarm concentration setting 2nd Alarm concentration setting CAL ZERO N2 SPAN CO2 Zero N2 calibration Span CO2 calibration CLr MAX Delete maximum alarmConcentration CLr STEL TWA...

Страница 10: ...button to save the value and move to the next step CAUTION Ensure that the high alarm setpoint must be greater than low alarm setpoint Alarm setpoints may vary by a country states and company policy Unless specified in your company s safety instruction use the preset alarm setpoints Before the alarm adjustment check with your safety manager or authorized dealer Ensure Standard Factory alarm set po...

Страница 11: ...PM150 CO2 Portable Gas Monitor REV 1 0 34 2900 0211 5 11 P a g e 7 Menu Screens 7 1 Menu Flow Chart Figure 7 1 ...

Страница 12: ...l CLr STEL TWA is displayed 2 Press the power button to delete the TWAand STEL values 3 After the successful activation SUC with V mark is displayed 7 4 Adjust Unit To change the unit ppm or Vol in the detector follow below steps 1 Press the Function key until UnIt is displayed and power button to enter the mode 2 Press the Function key to select a unit ppm or Vol and power button to save it 3 Aft...

Страница 13: ... displayed 2 Press the power button for three seconds In the SELF display press the power button for three seconds to activate the self test While it is activated the detector will test LED beeping vibration flash memory and sensor After the successful test V mark is displayed If the test fails FA with X mark is displayed 3 If the self test fails the Error message appears ...

Страница 14: ...alibration If the calibration continues to fail contact an authorized distributor or AimSafety 8 1 Calibration Gas Values Gas Type Zero N Span CO Concentration 99 99 Vol 20 000 ppm 2 Vol 8 2 Zero N2 Calibration 1 Assemble the calibration equipment with zero gas cylinder See Figure 8 1 2 In the CAL mode press the Function key to move to the ZErO calibration 3 Attach the calibration cap to the detec...

Страница 15: ...three 3 seconds to start the span calibration Once the countdown timer has started open the valve on the regulator to start the flow of gas 5 After 90 seconds when the span calibration is successful success message V appears If span calibration fails fail message X appears Note Press the power button during the 90 second countdown to abort the span calibration CAUTION Do not change the calibration...

Страница 16: ... Max 400 ppm 0 04 Vol 49 999 ppm 5 Vol Gas Concentration This allows a user to enter amend correct concentration of the gas cylinder Min Max 400 ppm 0 04 Vol 49 999 ppm 5 Vol Calibration Interval day The calibration reminder can be adjusted from 0 N A 365 days Default setting is N A Bump Interval Days The Bump test reminder can be adjusted from 0 N A 365 days Default setting is N A Self Interval D...

Страница 17: ... turn on the cylinder valve 5 The zero N2 and span CO2 calibrations each take 90 seconds 9 1 4 Log The most recent 30 logs will be stored in the device are automatically erased one by one the earliest logs when new events occur Thereare two types of logs Event Log and Event Data log available to download Choose the log and click Download button The logfiles will be downloadednamed with the unit s ...

Страница 18: ...rade U Open the firmware upgrade window to start upgrading process Self Test S Automatic self diagnostics of the unit Test order LED Beep Vibration Motor Flash Sensor End Factory Default F Reset original settings and specifications Time Write T To set the local time based on device settings Power OFF P Turn off the device IAP Version Read I Now Time When the Now Time button is clicked the PC time ...

Страница 19: ... Replacement of components can impact product performance Repairs should only be performed by authorized personal Only use replacement parts supplied by AimSafety A full calibration is required after sensor replacement Replacement Sensor 70 2900 0540 6 Replacement Sensor Filter pk 20 70 2900 0554 0 Replacement Battery 70 2900 0532 0 Replacement Front Cover 70 2900 0533 7 Mini USB Charger 70 2900 0...

Страница 20: ...d power up the device 6 After power up is complete and the sensor readings have stabilized for at least three 3 minutes perform Zero N2 99 9 vol calibration and Span CO2 calibration Figure 10 3 10 2 Battery Replacement 1 Move to a fresh air environment and power off the detector 2 Remove the back case by unscrewing the six 6 screws 3 Remove the two 2 screws on the PCB 4 Carefully unplug the batter...

Страница 21: ...tor 5V 1 2A While charging the backlight illuminates After the charging is complete the backlight turns off and a full battery symbol is displayed With a low battery the alarm is activated every three minutes 12 Appendix A Table of Figures Figure 3 1 5 Figure 7 1 11 Figure 8 1 14 Figure 9 1 16 Figure 10 1 19 Figure 10 2 20 Figure 10 3 20 Figure 11 1 21 Note The device can charge during communicati...

Страница 22: ...sing out of or related to the use of said gas monitor Manufacturer or its agent s liability shall be limited to replacement or repair as set forth above Buyer s sole and exclusive remedies are return of the goods and repayment of the price or repair and replacement of non conforming goods or parts Warranty Procedure Contact the local AimSafety authorized reseller or AimSafety Technical Support to ...

Отзывы: