PM150-CO2 Portable Gas Monitor
REV – 1.0
[34-2900-0211-5]
16
|
P a g e
9
Software Manager
9.1
Software Overview
Figure 9-1
•
Sensor Type
– The current sensor type in the device
(CO₂,
cannot be changed)
•
Serial Number
– PM150-CO
2
serial number
•
Firmware Version
– Current firmware version of the unit (can change by upgrading)
•
User ID/Message
– The User ID can be used to add a custom message
•
Unit
– Adjust measuring unit to ppm or %Vol
•
Stealth
– Disable the alarm, buzzer, and LED
•
Calibration Use
– Disable/Enable the calibration access
•
Low Alarm & High Alarm
– The 1
st
and 2
nd
alarm set points (Min/Max: 400 ppm (0.04% Vol)) - 49,999
capitalize (5% Vol)
•
STEL Alarm & TWA Alarm
– Short Term Exposure Limit and Time Weighted Average level of
concentration of
CO₂
(Min/Max: 400 ppm (0.04% Vol) - 49,999 ppm (5% Vol)
•
Gas Concentration
– This allows a user to enter/amend correct concentration of the gas cylinder
(Min/Max: 400 ppm (0.04% Vol) - 49,999 ppm (5% Vol)
•
Calibration Interval (day)
– The calibration reminder can be adjusted from 0 (N/A) – 365 days (Default
setting is N/A)
•
Bump Interval
(Days)
– The Bump test reminder can be adjusted from 0 (N/A) – 365 days (Default setting
is N/A)
•
Self-Interval
(Days)
– The Self-test reminder can be adjusted from 0 (N/A) – 365 days (Default setting is
N/A)
Note:
-
When the software is opened, the fields are grayed out and before it can be used, the “OPEN” button must
be clicked.
-
Without clicking the “Write” button, configured and customized settings will not be applied, and neither be
saved.
-
If the USB connection is successful, the “Success” icon appears. If the connection fails, reconnect
the USB cable or check the device manager to see the connection status.
Read
Write
Calibration
Log
Firmware