Getting Started with FeverWarn l Models FW1100A and FW1100B 15
MachineSense • Sales: +1-443-457-1165 • [email protected] • www.feverwarn.com
Figure 32
Figure 33
Figure 34
Using the FeverWarn App
Calibration
Your FeverWarn scanner was calibrated at the factory to scan and
record accurate temperatures in Fahrenheit or Celsius. However, you
do have the option to recalibrate the sensor. This process requires
a handheld medical thermometer, and you should have a second
person helping you.
1. To begin, tap
START CALIBRATION.
2. A pop-up will appear to verify you want to proceed with the
calibration. Triggering a calibration alert will disable thermal
detection. It will be enabled again after completing the calibration
process. Faulty calibration will cause the scanner to behave
incorrectly. Do not start the calibration process until you have
read and understood the process. Tap
YES to proceed (Figure 32).
Wait while the alert is triggered
(Figure 33).
3. Wait for the device to be ready to calibrate. You will see a message
“Calibration started successfully”
(Figure 34).
(continued on next page)
QC-FW005-051821