2.7.2 DETAILED SETTING
Selectable operating functions are as shown in the table below.
Arithmetic
Addition
K1X1 + K2X2 + A1
Multiplication
(K1X1+A1) (K2X2+A2) + A3
Division
(K1X1+A1) / (K2X2+A2) + A3
Logical
AND
X1 X2
OR
X1 X2
NOT
¬ X1
XOR
X1 ^ X2
Math
Square root
Square root extraction K1 √ X1
Power
Power X1
A1
Accumulation
Analog accumulation, Pulse accumulation
Peak hold
Peak hold (max)
Maximum value hold
Peak hold (min)
Minimum value hold
Filter
First order lag
Time constant is a response time for a step input (0 to
100%) to reach 63%.
Moving average
Multiple samples of input data are averaged.
F value calculation
Typically used to calculate the sterilization or disinfection
time in predefined conditions
When you choose a type of operating functions, function blocks appear at the bottom of the window.
Addition / Subtraction
Multiplication
Division
AND
OR
NOT
XOR
Square root
>
>