21
________________________________________________________________________________
A1. Specifications
A1.1 System Classifications
Classification according to type of protection against electric shock.
CLASS 1 EQUIPMENT
Classification according to the degree of protection against electric shock.
TYPE B EQUIPMENT
Classification according to degree of protection against harmful ingress of
water.
NO CLASSIFICATION
Classification according to method(s) of sterilisation or disinfection
recommended by the manufacturer.
NO CLASSIFICATION
Classification according to the degree of application in the presence of a
flammable anaesthetic mixture with air or with oxygen or with nitrous oxide.
NO CLASSIFICATION
Classification according to the mode of operation.
CONTINUOUS OPERATION
The Equipment is also classified as PORTABLE.
A1.2 System Specifications
The specifications given in the present section are the general specifications
satisfied by all Luminette Q Systems. For the Test Specifications of the user's
own system he/she is referred to the separate Test Sheets supplied with this
manual.
A1.2.1 Luminette Q Parameters
Light Source
: Passively Q-switched, flashlamp-pumped Nd:YAG.
Internal Coolant
: Deionised Water.
A1.2.2 Output Parameters
Wavelength
: 1064nm, 532nm
Pulse Train Energy
: Up to 1600mJ
Repetition Rate
: 0.63Hz, 1.25Hz, 2.5Hz, 5.0Hz
Spotsizes
: 3mm diameter (Accurate to 5%)
: 5mm diameter (Accurate to 5%)
Energy Densities
: 0
–
22.6 J/cm
2
*