Lynton Luminette Q Скачать руководство пользователя страница 23

19 

________________________________________________________________________________ 

4.  Cleaning and Maintenance 

 

4.1  Cleaning 

 
The Luminette Q Laser System cabinet, control panel and handpiece may be 

cleaned  on  a  routine  basis  by  simply  wiping  down  with  a  damp  cloth.  The 
spotsize spacer attachments should be kept clean by wiping them with lens 

tissue moistened with an organic solvent such as ethanol or methanol. It is 

recommended that this is done between patient treatments.  
 

4.2  Maintenance Schedule 

 

    Daily 

- Check for water leaks external to the cabinet . 
- Check for damage to the spotsize spacers. 

 Every 1,000,000 shots  -  Change Flashlamp. 

- Check cooling water level. Top up if required 

- Check internal water connections for tightness.  

4.3  Maintenance Procedures 

 

Other  than  the  daily  checks  it  is  not  recommended  that  the  general  USER 

attempt  any  of  the  maintenance  procedures  listed  in  the  previous  section.  
Only competent and authorised personnel (as described in the WARNING box 

above) should perform the maintenance procedures, which are detailed fully in 
the  Service  Manual.  This  and  any  additional  relevant  material  (e.g.  Circuit 

Diagrams,  Mechanical  Design  Drawings  etc.)  can  be  made  available  after 

consultation with Lynton Lasers Ltd.  Service Contracts are also available. 
 

4.3.1  Topping up the Cooling Water 
 

If there is insufficient deionised water in the cooling system the pump will not 

start up when the keyswitch is turned on and it will be impossible to fire the 
flashlamp in the Luminette Q System. To check that insufficient water is the 

real cause of the problem, look at the column of red interlock LEDs on the 
back  panel.  If  No.3  is  lit  when  the  keyswitch  is  turned  on  then  insufficient 

water in the cooling system is definitely the problem (No.1 will also be lit). 

Simply unscrew the black Reservoir Filler Cap from the Reservoir Filling Point 
on the back panel of the Luminette Q System and top up the reservoir with 

250ml of deionised water using the Filling Bottle shipped with the Luminette 

CAUTION    :    Only  personnel  from  Lynton  Lasers  Ltd,  or  from  companies 

whose personnel have been trained by Lynton Lasers Ltd, should perform 

maintenance procedures. The schedule outlined below does not present any 
increase  in  light  hazard  unless  specifically  stated.  The  schedule  must  be 

adhered to.  Failure to do so does not lead to an increased light/electrical 

hazard but could result in serious damage to the equipment. 

Содержание Luminette Q

Страница 1: ......

Страница 2: ......

Страница 3: ...he Luminette Q Laser System 12 3 5 Changing Handpieces 15 3 6 Interlock Warning Screen and LEDs 16 4 CLEANING AND MAINTENANCE 19 4 1 Cleaning 19 4 2 Maintenance Schedule 19 4 3 Maintenance Procedures...

Страница 4: ...A1 3 Available Options and Accessories 22 A1 4 Labelling 23...

Страница 5: ...the Luminette range of IPL laser systems the standard Luminette System the Luminette Advance system and the recently introduced Luminette Q system All of these models look similar from a physical and...

Страница 6: ...ttons Pre set default parameters are included for the different Skin Types all determined through clinical experience The output of the Luminette Q Laser System comprises of passively Q switched laser...

Страница 7: ...onent Users are particularly recommended to read the accompanying Training Manual before using the Luminette Q System for its intended purposes in cosmetic dermatology CAUTION The Luminette Q Laser Sy...

Страница 8: ...detecting equipment should be used The radiation classed as completely invisible may interact with any surface on which it impinges to produce visible light The user is warned that even this light may...

Страница 9: ...chable Handpiece properly connected it does not present the operator with any electrical hazard When Detachable Handpieces are being changed over there are no hazardous voltages on any of the electric...

Страница 10: ......

Страница 11: ...ected to ensure that it has not been unreasonably damaged during transit If the crate has been damaged it is recommended that the supplier is contacted immediately It will then be recommended that pho...

Страница 12: ...Section 4 3 2 of Chapter 4 Cleaning Maintenance Because the internal cooling system requires an air vent a small hole in the pipe of the reservoir filling point to operate properly it is likely that...

Страница 13: ...ressed The key cannot be removed in this position Turn the key switch counter clockwise to turn the Luminette Q System off again at any time Remove the key and store in a safe location when the Lumine...

Страница 14: ...e Q System is shown below 6 Fused Mains Inlet Pulling out the detachable mains lead from this socket removes all power from the Luminette Q System Alternatively the mains lead can unplugged from the w...

Страница 15: ...ividual LEDs comes later 10 Reservoir Filling Point The internal cooling system can be topped up at this point using the pipe provided The pipe is normally fitted with a two position cap In normal use...

Страница 16: ...ns performed at any time will remove the possibility of the emission of light i Turning the key switch counter clockwise to the OFF position ii Pressing the Emergency Stop Button iii Switching off or...

Страница 17: ...ers can be selected and the system can be enabled for treatment This Standby screen displays the handpiece type currently connected to the Luminette Q system in the HP area of the screen As can be see...

Страница 18: ...ing this button increases the rate When the maximum Rate is reached the setting then reverts back to the minimum Rate At this point the User can fire the handpiece by pressing the Pushbutton on the Ha...

Страница 19: ...to remove a Detachable Handpiece and re connect another one To remove a Detachable Handpiece it is necessary to simultaneously press the two buttons on either side of the connector socket whilst at th...

Страница 20: ...le to fire the Flashlamp and the screen will change to INTERLOCK BROKEN Door Interlock Close the Door Press any Key If however there is a more serious Water Flow Water Level or Water Temperature fault...

Страница 21: ...hot Inhibits Flashing If the fault cannot be easily corrected contact Lynton Lasers Ltd or one of its designated agents stating which LED s is are lit If a handpiece reaches the end of it s recommende...

Страница 22: ...18 ________________________________________________________________________________...

Страница 23: ...rial e g Circuit Diagrams Mechanical Design Drawings etc can be made available after consultation with Lynton Lasers Ltd Service Contracts are also available 4 3 1 Topping up the Cooling Water If ther...

Страница 24: ...will light up as well as No 1 Remove the black Reservoir Filler Cap from the Reservoir Filling Point on the back panel of the Luminette Q System and fill up with deionised water using the Filling Bott...

Страница 25: ...able anaesthetic mixture with air or with oxygen or with nitrous oxide NO CLASSIFICATION Classification according to the mode of operation CONTINUOUS OPERATION The Equipment is also classified as PORT...

Страница 26: ...r a suitable pair of glasses or goggles for the Luminette Q Laser System contact Lynton Lasers Ltd A1 2 5 Electrical Parameters Operating Voltage Range 100 125 or 200 264 Volts AC 2 engineer selectabl...

Страница 27: ...ating Plate panel of the label B Panel B Rating Plate panel containing the rated mains voltage with the sign indicating that it is an AC voltage the input mains power the input mains frequency the Ele...

Страница 28: ...t the Luminette Q is a Class 4 Laser System Laser Radiation Starburst Hazard Label Label indicating the requirement for efficient recovery of the Luminette Q System at end of use according to the WEEE...

Страница 29: ...Lynton House Manor Lane Holmes Chapel Cheshire CW4 8AF United Kingdom Telephone 44 0 1477 536977 Fax 44 0 1477 536978 email Info lynton co uk...

Отзывы: