background image

ASSEMBLY OF HEADLIGHT SYSTEMS

4

Next, place the headband or sweatband on your head.  
You can make size adjustments on the headband by 
turning the knob at the rear of the headband until it fits 
comfortably.  The cross band can be adjusted for even 
more comfort.  Push out the tab and slide the band until 
the fit is correct, then replace the tab.  The sweatband 
should be pulled to a comfortable tension and closed 
using the velcro strip 

(Figure 4). 

 

To remove the module from the headband or 
sweatband, a small screwdriver is needed.  Insert the 
screwdriver into the slot at the top of the module 
receptacle.  Lift the handle of the screwdriver and the 
module will slide out. 

Caution: 
 
To plug or unplug the fiber optic cable from the 
light source, it is recommended that the light 
source power be turned off.  Be sure to hold the 
strain reliever near the end of the cable.  This 
part of the cable is reinforced to prevent damage 
to the glass fibers inside, and to prevent burns. 

Figure 3

SWEATBAND

HEADBAND

Fiber 
optic
cable

Cable 
Clips

Headlight
module

Joystick

Headlight
module

Joystick

Fiber 
optic cable

Cable 
Clips

Figure 4

HEADBAND

SWEATBAND

Cable 
Clip

Cable 
Clips

Adjusting knob

Velcro
strip

ADJUSTMENTS

Once the headlight module and fiber optic cable are attached to the headband, the headlight assembly is ready to use. 

 

The headlight module can be adjusted while wearing the headlight system by grasping the joystick.  The joystick can 
be removed for autoclaving or to prevent misaligning the headlight module when working in close quarters.  The iris 
adjustment varies the spot size from 20mm to 100mm (0.8 to 4 inches) at a 40 cm (16 inch) working distance 
(

Figure 1a

). 

 

The headlight system comes with two gown clips to relieve the weight of the cable from the headband.  The upper clip 
is attached to the surgical gown just below the shoulder (

Figure 5

).  Be sure to leave enough slack in the cable to 

allow for freedom of head movement.  If the surgeon needs to detach the cable from the light source, the lower gown 
clip is used to attach to the surgeon’s gown.  This eliminates the risk of dropping or tripping over the cable.

Содержание UltraLife

Страница 1: ...UltraLite Headlight Operation and Service Manual ENGLISH...

Страница 2: ...ng instructions This manual should also be available to the surgical team during a procedure Follow the instructions in the operating manuals of other manufacturer s equipment when they are use in con...

Страница 3: ...to curve gently around the shape of the headband All illustrations are representative and may not reflect your actual model and accessories HEADBAND properly supported Remove the cable from the module...

Страница 4: ...burns Figure 3 SWEATBAND HEADBAND Fiber optic cable Cable Clips Headlight module Joystick Headlight module Joystick Fiber optic cable Cable Clips Figure 4 HEADBAND SWEATBAND Cable Clip Cable Clips Ad...

Страница 5: ...ralite 2 bifurcated cable Ultralite module and extended linkage AX 1375 BIF SL Headband with Ultralite 2 bifurcated cable and Ultralite module AX 1385 BIF SL Sweatband with Ultralite 2 bifurcated cabl...

Страница 6: ...e the cables in sterilization trays for additional protection Any inadvertent cut or puncture to the silicone tube will render the cable unsafe It should be taken out of service immediately Cleaning M...

Страница 7: ...romptly reports such a defect to LUXTEC in writing 2 If requested by LUXTEC customer returns equipment to LUXTEC with shipping charges pre paid and 3 Upon inspection LUXTEC finds the cable to be defec...

Страница 8: ...s the Essential Requirements specified in Annex 1 of the Medical Device Directive MDD 99 Hartwell Street West Boylston MA 01583 USA Telephone 1 800 325 8966 US only or 1 508 835 9700 Facsimile 1 508 8...

Отзывы: