CARE AND MAINTENANCE

TROUBLESHOOTING

For additional technical assistance, please contact the Luxtec Service Department at 1-800-325-8966 
(USA only) or +1-508-835-9700, Fax +1-508-835-9976 or email: [email protected]  

6

All Headlight Systems: 

To ensure long life for the headband, cable, and 
modules, they should be stored in the duffel bag or 
case provided.  Fiber optic cables are fine, high 
quality optical instruments.  You can prolong the 
useful life by following a few guidelines: 

• 

Avoid stretching your cable, forming 
configurations involving sharp angles or kinks, 
or contact with sharp or pointed objects.  The 
internal light fibers are made of glass, a material 
which breaks under stress.  Fiber breakage will 
result in diminished light output. 

• 

Utilize the gown clips (If provided) to keep the 
headlight cable secured to the surgeon’s gown 
for greater comfort and maneuverability. (This 
does not include endoscopic use) 

• 

Do not use your fiber optic cable with any 
alterations to it’s original design or fabrication. 
Bundle size (aperture) of the fiber optic cable 
should be matched to the aperture of the 
instrument to obtain maximum light output. 

• 

NOTE: A larger aperture fiber optic cable will 
not increase light output  of a smaller aperture 
instrument. 

• 

Keep the optical faces from contacting the floor 
or other hard surfaces. The resulting scratches 
will diminish light output.  Store the cables in 
sterilization trays for additional protection. 

• 

Any inadvertent cut or puncture to the silicone 
tube will render the cable unsafe.  It should be 
taken out of service immediately. 

 

Cleaning Modules:

 

The headlight module may be wiped down with 
alcohol.   The lens should be cleaned only with lens 
tissue available at any camera store.  Follow the 
package directions.   
Follow all applicable bloodborne pathogen cleaning 
procedures as indicated by OSHA and / or your 
hospital requirements when cleaning, disinfecting, 
and sterilizing instruments and accessories. 

Cleaning Joystick (only): 

The joystick can be removed from the module for 
autoclaving.

 

 
Cleaning Sweatband: 

The sweatband may be cleaned using standard wash 
cycles in warm water and detergent. Air dry. 
 

Cleaning Headband: 

The headband may be cleaned by an alcohol 
wipedown or washed with lukewarm water and mild 
detergent.  Rinse thoroughly with water. 
 

Cleaning Cable: 

Luxtec suggests that the use of alcohol wipes to 
clean the UltraLite and Sweatband headlight cables. 
 

Disinfecting Cable: 
Soaking the UltraLite 2 cables in any sterilizing 
disinfectant should be avoided.

  There is a lens 

assembly at the light source end, which is not sealed 
and could allow moisture to penetrate behind the 
lensing. 
 

Sterilizing Cable: 

The UltraLite fiber optic cable has not been 
specifically designed to withstand sterilization.  
There is a lens assembly at the light source end, 
which is not sealed and could allow moisture to 
penetrate behind the lensing. 
 

WARNING: Do not autoclave the headband, 
cables, sweatband or headlight modules.

 

Содержание UltraLife

Страница 1: ...UltraLite Headlight Operation and Service Manual ENGLISH...

Страница 2: ...ng instructions This manual should also be available to the surgical team during a procedure Follow the instructions in the operating manuals of other manufacturer s equipment when they are use in con...

Страница 3: ...to curve gently around the shape of the headband All illustrations are representative and may not reflect your actual model and accessories HEADBAND properly supported Remove the cable from the module...

Страница 4: ...burns Figure 3 SWEATBAND HEADBAND Fiber optic cable Cable Clips Headlight module Joystick Headlight module Joystick Fiber optic cable Cable Clips Figure 4 HEADBAND SWEATBAND Cable Clip Cable Clips Ad...

Страница 5: ...ralite 2 bifurcated cable Ultralite module and extended linkage AX 1375 BIF SL Headband with Ultralite 2 bifurcated cable and Ultralite module AX 1385 BIF SL Sweatband with Ultralite 2 bifurcated cabl...

Страница 6: ...e the cables in sterilization trays for additional protection Any inadvertent cut or puncture to the silicone tube will render the cable unsafe It should be taken out of service immediately Cleaning M...

Страница 7: ...romptly reports such a defect to LUXTEC in writing 2 If requested by LUXTEC customer returns equipment to LUXTEC with shipping charges pre paid and 3 Upon inspection LUXTEC finds the cable to be defec...

Страница 8: ...s the Essential Requirements specified in Annex 1 of the Medical Device Directive MDD 99 Hartwell Street West Boylston MA 01583 USA Telephone 1 800 325 8966 US only or 1 508 835 9700 Facsimile 1 508 8...

Отзывы: