26
0698-701 Relay module - 04/2021
en
2. Trouble shooting
2.1 Fault report
Operating mode Fault report
Cause
Remedy
General
Communication with
another module not
possible.
• The connection
between the modules
is interrupted
• Check bus line
CHECK BUS!
• Program error in the
system
• Restart the system by pressing
"Reset" on the operating unit
• Module damaged
• Replace the module
The converted
part. vol. of
...........
is too large for display.
Excess will be deleted.
• Smaller measurement
unit was selected.
The complete partial
volume can no longer
be displayed after the
conversion.
The complete partial volume
is calculated from the sum of
the displayed volume and the
maximum displayable volume.
Example:
Displayed volume: 104.3 L
Max. displayable volume
99999,9 L
Complete partial volume:
99999,9 + 104,3 = 100104,2
Liters
The converted total
volume of
.........
is too large for display.
Excess will be deleted.
• Smaller measurement
unit was selected
The complete total volume is
calculated from the sum of
the displayed volume and the
maximum displayable volume.
Example:
Displayed volume: 104.3 L
Max. displayable volume:
99999.9 L
Complete partial volume:
99999,9 + 104,3 = 100104,2
Liters
The converted partial
and total volumes of
..........
are too large for
display. Excess will be
deleted.
• Smaller measurement
unit was selected
see above
An unknown error has
occured. Please check
the settings in the
operating unit.
• During the start of the
system, liquid flows
through the basic flow
meter
Prevent a volume flow through
the basic flow meter during the
start of the system