32 General Operation
Setting the power
The maximum available treatment power varies depending on the type of
delivery device used, ranging from 1800mW, nominal for slit lamps, to
2500 mW, nominal for Acculite endoprobes.
To set the power from the console, press the Treat Power
(increase) and
(decrease) selectors until the desired setting displays. To set the power from the
remote control, press the
(increase) and
(decrease) selectors in the
mW
(power) field until the desired setting displays. The system will beep
when the maximum available power is selected.
If you are using the PowerEase footswitch, you can also adjust the laser power
by pressing the switches on the inside left (decrease power) and right (increase
power) of the footswitch housing.
WARNING -
Always inspect the optics and delivery device as describedin
“Connection Instructions” to ensure that the device is clean and undamaged.
Never use a device that appears damaged or dirty. Doing so may dangerously
reduce the delivered power and result in unintended, adverse tissue effects.
Setting the exposure time (pulse duration)
The pulse duration setting determines the duration of each laser exposure.
Exposure time settings range from 0.01 to 3.0 seconds. If you release the
footswitch before the selected exposure time has elapsed, laser exposure is
interrupted.
To set the pulse duration, press the up or down pulse duration selectors on
either the laser console or remote control until the desired value displays.
Содержание Novus Spectra
Страница 1: ...Novus Spectra Solid State 532 nm Green Laser Operator Manual...
Страница 6: ...6 Contents Novus Spectra 0642 543 01 Rev A 12 02...
Страница 8: ...Novus Spectra 0642 543 01 Rev A 12 02 8 General Operation...
Страница 40: ...Novus Spectra 0642 543 01 Rev A 12 02 40 General Operation...
Страница 42: ...Novus Spectra 0642 543 01 Rev A 12 02 42 Maintenance...
Страница 56: ...Novus Spectra 0642 543 01 Rev A 12 02 56 Safety and Regulatory...
Страница 72: ...Novus Spectra 0642 543 01 Rev A 12 02 72 Indications for Use...
Страница 110: ...Novus Spectra 0642 543 01 Rev A 12 02 110 Indications for Use...