Novus Spectra
0642-543-01, Rev A, 12/02
18
General Operation
Connecting the delivery device(s)
WARNING -
To avoid possible damage to the laser, use only qualified,
compatible Lumenis delivery devices. Using other than Lumenis delivery
devices may jeopardize safe operation or damage the laser and will void your
Lumenis warranty or service contract.
WARNING -
When using a fiber optic delivery device, always inspect the
fiber optic cable to ensure that it has not been kinked, punctured, fractured, or
otherwise damaged. The fiber optic cable may be damaged if stepped on, pulled,
left lying in a vulnerable position, kinked, or tightly coiled. Do not clamp the
cable with a hemostat or other instruments. If sterile tape is used, always
remove the tape before lifting the cable. A damaged fiber optic cable may cause
accidental laser exposure or injury to the treatment room personnel or
patient, and/or fire in the treatment room.
WARNING -
Carefully inspect any delivery system sterile packaging to ensure
that it has not been torn or punctured. If there is any damage to the sterile
packaging, do not use the delivery system
WARNING -
If an emulation plug is used, the operator shall assume all
responsibility for ensuring that the proper eye safety filters are used.
A variety of delivery devices are compatible with this laser. To connect a
delivery device, attach the delivery device laser connector to the fiber port and
delivery device receptacle (ACC), as appropriate, on the front of the laser
console. Refer to the appropriate delivery device operator manual for detailed
connection and operating instructions.
When a delivery device is properly connected, the corresponding icon displays
on the remote control.
represents a LaserLink
™
or slit lamp
represents a probe or handpiece
represents an LIO
™
If the delivery device is not properly connected, an E300 error code will
display on the laser console, and the delivery device icon willnot illuminate on
the remote control.Verify that the device is properly connected to the
appropriate receptacles before turning the laser on. If the delivery device is not
compatible with the Novus Spectra laser system, an E301 error code will
display.
Содержание Novus Spectra
Страница 1: ...Novus Spectra Solid State 532 nm Green Laser Operator Manual...
Страница 6: ...6 Contents Novus Spectra 0642 543 01 Rev A 12 02...
Страница 8: ...Novus Spectra 0642 543 01 Rev A 12 02 8 General Operation...
Страница 40: ...Novus Spectra 0642 543 01 Rev A 12 02 40 General Operation...
Страница 42: ...Novus Spectra 0642 543 01 Rev A 12 02 42 Maintenance...
Страница 56: ...Novus Spectra 0642 543 01 Rev A 12 02 56 Safety and Regulatory...
Страница 72: ...Novus Spectra 0642 543 01 Rev A 12 02 72 Indications for Use...
Страница 110: ...Novus Spectra 0642 543 01 Rev A 12 02 110 Indications for Use...