Operating Manual V25/09.2019
Ventus / Ventus-X / V200A
Chapter 20 Appendix
72
Sensor Status:
Each register hoIds 4 sensor status coded with 4 bits per status. The sequence defined in the
table above is to understand as from most significant half byte to least significant half byte. Most
of the sensors have two status values, one for the sensor itself and the current measurement
value, another one for the buffer, from which average, min. And max values are evaluated.
Assignment of Status Information to Status Register
Register
Byte
Half-
Byte
Status
Sensor Status 1
High
High
Temperature Buffer
Low
Temperature
Low
High
Air Pressure Buffer
Low
Air Pressure
Sensor Status 2
High
High
Wind Buffer
Low
Wind
Low
High
0
Low
0
Example Sensor Status 1
:
Temperature buffer status, temperature status, air pressure buffer status, air pressure status
High Byte
Low Byte
High
Low
High
Low
Temp
e
ra
tu
r
e
-B
u
ff
e
r
Temp
e
ra
tu
r
e
A
ir
P
re
s
s
u
re
B
u
ff
e
r
A
ir
P
re
s
s
u
re
5
3
0
7
The example values above (for illustration only, the given combination will not occur in reality)
are combined to the register value 0x5307 = 21255.
The single status are retrieved from the register as integer part of
Status 1
= register / 4096
Status 2
= ( register / 256 ) AND 0x000F
Status 3
= ( register / 16 ) AND 0x000F
Status 4
= register AND 0x000F
Following table shows the status coding:
Содержание 8371.UA01
Страница 1: ......
Страница 2: ...Operating Manual V25 09 2019 Ventus Ventus X V200A 2 ...
Страница 29: ...Operating Manual V25 09 2019 Ventus Ventus X V200A Chapter 12 Configuration and Test 29 Figure 15 Channel list ...
Страница 30: ...Operating Manual V25 09 2019 Ventus Ventus X V200A Chapter 12 Configuration and Test 30 Figure 16 Data query ...
Страница 65: ...Operating Manual V25 09 2019 Ventus Ventus X V200A Chapter 20 Appendix 65 0XM11 CR LF ...
Страница 76: ......