Chapter 12 Configuration and Test
26
12.2.6 Heating Parameters
Figure 12: Heating Parameters
Setpoint temperature:
Temperature to which the
Ventus
is heated
default: 15°C
Heater mode:
off:
heating is off
auto:
automated control to setpoint temperature
testmode:
Ventus
is heated to test setpoint temperature
ctrl high:
heating off if control line on high,
otherwise: automatic control
ctrl low:
heating off if control line on low
otherwise: automatic control
default: auto
Capacity:
full:
full heating power
alternating:
upper plate and lower plate alternatingly
default: full
Disable if overvoltage:
the heating is disabled if overvoltage (28V) is detected
default: yes
Disable if undervoltage:
the heating is disabled if undervoltage (20V) is detected
default: no
12.2.7 Wind Parameters
Figure 13: Wind Parameters
Min. speed:
starting speed of the wind sensor in m/s
default: 0,1m/s
Counts of measurements:
Number of measured values used for determining the statistical values
of average, minum value and maximum value
default: 60
Direction Offset:
Correction of the wind direction
It is possible to have an offset added to the measured wind direction,
in order to convert e.g. 0° (North) into 180° (South)
default: 0°
Measurement rate:
Update interval, adjustable in 250ms
, 1 … 10 s
default: 10s
12.2.8 Pressure Parameters
Pressure offset:
Absolute offset (for on-site adjustment) on the measured value
Содержание 8371.UA01
Страница 1: ......
Страница 2: ...Operating Manual V25 09 2019 Ventus Ventus X V200A 2 ...
Страница 29: ...Operating Manual V25 09 2019 Ventus Ventus X V200A Chapter 12 Configuration and Test 29 Figure 15 Channel list ...
Страница 30: ...Operating Manual V25 09 2019 Ventus Ventus X V200A Chapter 12 Configuration and Test 30 Figure 16 Data query ...
Страница 65: ...Operating Manual V25 09 2019 Ventus Ventus X V200A Chapter 20 Appendix 65 0XM11 CR LF ...
Страница 76: ......