Foggy
no
Small amount
Massive
Multiple of attenuation
0dB
5...10dB
10...20dB
Percent attenuation
0%
50~67%
67~90%
Increase the range multiple
No need to increase
1x range
3 times the range
With steam
no
Small amount
Massive
Multiple of attenuation
0dB
5...10dB
10...20dB
Percent attenuation
0%
50~67%
67~90%
Increase the range multiple
No need to increase
1x range
3 times the range
Probe and media surface
temperature difference
≤20℃
≤40℃
≤80℃
Multiple of attenuation
0dB
5...10dB
10...20dB
Percent attenuation
0%
50~67%
67~90%
Increase the range multiple
No need to increase
1x range
3 times the range
Signal attenuation calculation method. If there are multiple situations in the field, it is necessary to add
up these several situations to calculate the amount of signal attenuation.
Total minimum 15dB, maximum 40dB.
▲ It is recommended to use a plastic flange to connect the sensor during installation
4.2.3 bottom thread installation
In this case, if the actual maximum measurement range is 5 meters, you need to use the ultrasonic
level meter with a range of 50 meters.
Temperature difference between probe and medium surface ≤40 ℃ 5 ... 10dB
With a small amount of feed 5 ... 10dB
With a small amount of steam 5 ... 20dB