Model 3030 & 3030E Series
Technical Manual
Section 7
Ludlum Measurements, Inc.
Page 35
November 2017
Cn - Count
Performs various count functions. The count time is determined by the
positing of the COUNT TIME SWITCH on the front panel unless
otherwise noted. See command RR for the format of the count results.
1 = Start a count.
2 = Stop a count in progress.
3 = Clear alarms.
4 = Start a count and automatically return the results.
5 = Start a one-minute count and automatically return the results.
Dn – Display Mode
The display can be set to show count data as raw scaler counters, counts per
minute (CPM), and disintegrations per minute (DPM).
0 = Set display mode to scaler.
1 = Set display mode to CPM.
2 = Set display Mode to DPM.
3 = Return display mode (SCA, CPM, or DPM).
F – Firmware
Returns a string containing the firmware version number.
Ln – Logging Mode
Sets how samples are logged to internal memory.
0 = Turn logging OFF.
1 = Turn Logging ON.
2 = Log only QC checks.
3 = Return logging mode (OFF, ON, or QC).
Pn
Sets printer mode On/Off.
0 = Turn printing OFF.
1 = Turn printing ON.
2 = Return printing mode (OFF, ON).
Qx
Sets quick power-on mode. The Model 3030 may be configured to display
certain parameters during power-up.
Содержание 3030
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 56: ...Model3030 3030ESeries TechnicalManual Section7 Ludlum Measurements Inc Page 44 November 2017 ...
Страница 57: ...Model3030 3030ESeries TechnicalManual Section7 Ludlum Measurements Inc Page 45 November 2017 ...
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......