Model 3030 & 3030E Series
Technical Manual
Section 7
Ludlum Measurements, Inc.
Page 29
November 2017
10.
Press the COUNT button to exit the QC check mode.
HV Plateau
A high-voltage (HV) plateau is normally run at calibration time in order to
set the proper HV and to determine the operating efficiencies. This routine
is provided to automate the plateau procedure and simplify the calculations.
This screen allows you to enter the starting and ending value, the voltage
increment, and the source size. The background count is always one minute.
The data may be saved in a text file or printed to a printer.
The plateau order can be changed from Alpha, Beta, Background to Beta,
Alpha, Background.
STEPS FOR PERFORMING A HIGH VOLTAGE PLATEAU
1.
Turn the Model 3030 on and connect to a computer by way of a
serial port with supplied cable.
2.
Run the Model 3030 control software. Software will automatically
detect the instrument and download all parameters.
3.
When all the parameters have been downloaded, click on the tab
marked “HV Plateau” at the top of the window.
Содержание 3030
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 56: ...Model3030 3030ESeries TechnicalManual Section7 Ludlum Measurements Inc Page 44 November 2017 ...
Страница 57: ...Model3030 3030ESeries TechnicalManual Section7 Ludlum Measurements Inc Page 45 November 2017 ...
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......