44
Technical data
EN
Max. CPAP pressure
in the event of a fault
to EN ISO 17510 -1
< 36 hPa
Flow rate at max.
speed at:
20 hPa
16 hPa
12 hPa
8 hPa
4 hPa
0 hPa
Tolerance
125 l/min
145 l/min
165 l/min
180 l/min
195 l/min
210 l/min
±15 l/min
115 l/min
135 l/min
150 l/min
165 l/min
180 l/min
190 l/min
±15 l/min
Heating of respiratory
air
2,5 °C (as per HMV - [Heilmit-
telverordnung – German regu-
lations governing
pharmaceutical products])
Depends on heating level
Short-term pressure
consistency measured
to EN ISO 17510-1 at:
20 hPa
16 hPa
13 hPa
10 hPa
7 hPa
4 hPa
Δ
p = 0.6 hPa
Δ
p = 0.5 hPa
Δ
p = 0.4 hPa
Δ
p = 0.4 hPa
Δ
p = 0.4 hPa
Δ
p = 0.3 hPa
Δ
p = 0.6 hPa
Δ
p = 0.5 hPa
Δ
p = 0.4 hPa
Δ
p = 0.4 hPa
Δ
p = 0.4 hPa
Δ
p = 0.3 hPa
SOMNO
vent
auto-S/ST
SOMNO
vent
auto-S/ST with
SOMNO
click
300