30
Hygiene treatment
EN
6.2 Breathing tube
Caution!
The breathing tube may only be used again after cleaning once it is completely dry.
If moisture gets into the therapy device, this may lead to damage to the device and
thus to a risk to users and patients.
Clean breathing tube
1. Take the breathing tube off the therapy device and off the exhalation system.
2. Seal both sides of the pressure measurement tube with a sealing plug.
3. Clean the creased tube and the click adapter with a little detergent in hot water so that
there are no residues. Rinse the inside of the tube thoroughly in the process.
4. Then flush the creased tube thoroughly inside and outside with clean, hot water.
5. Shake out the breathing tube thoroughly.
6. Hang up the breathing tube and allow it to drip dry.
7. Dry the breathing tube with the aid of the therapy device.
Dry breathing tube
You can dry the breathing tube completely with the aid of the therapy device. To do so,
proceed as follows.
Status:
therapy device is switched off.
1. Remove the respiratory air humidifier if required.
2. Remove the plugs from the pressure measurement tube.
3. If water gets into the pressure measurement tube acci-
dentally, plug the red drying adapter supplied into the
device outlet port.
As required
Disinfect therapy device
Disinfect SOMNO
click
300
See associated
instructions for use
On change of
patient
Hygiene treatment if required
Change particulate filter of bacteria filter.
See associated
instructions for use