Hygiene treatment
33
EN
Ensure that no liquids penetrate the therapy device. Never immerse the therapy de-
vice in disinfectants or other liquids, otherwise this may lead to damage to the ther-
apy device and thus to a hazard to users and patients.
Caution!
Ensure that no liquids penetrate the therapy device via the pressure measurement
connection, otherwise there may be malfunctions.
The housing and the power cord of the therapy device are cleaned simply by wiping with
disinfectant.
Breathing tube
We recommend gigasept
®
FF for disinfecting by immersion.
For disinfecting, proceed exactly as for cleaning.
Caution!
The breathing tube may not be heated to over 70 °C. Sterilization is not permitted.
Accessories
For disinfecting/sterilizing the exhalation system, the breathing mask and the
SOMNO
click
300 respiratory air humidifier, see the section entitled “Disinfect and sterilize”
in the corresponding instructions for use.
6.7 Change of patient
Caution!
• If the therapy device has been used without a bacteria filter, the therapy device
must be subjected to a hygiene treatment before it can be used for another
patient. This must be performed by the manufacturer or by an authorized
specialist dealer.
• The procedure for the hygiene treatment is described on the servicing sheet and
in the Servicing and Repair Instructions for the therapy device.