56
Servicing
EN
9. Servicing
1. Use the PC software to calibrate the pressure sensor for flow measurement every 2 years.
The procedure is described in the User Manual for the PC software.
2. To read off the next calibration date: Call up the
Erase data
page by briefly pressing the
button on the device several times.
Erase data
and
Next calibration
appear in the
display.
3. If you are not performing calibration yourself using the PC software: send the device to
the manufacturer or specialist staff expressly authorized by the manufacturer every 2
years to have servicing performed.
4. Perform a function check (see “7. Function check” on page 52).
9.1 Storage
Store the device under the specified ambient conditions (see “11.1 Specifications” on
page 61).
Remove the batteries/rechargeable batteries from the device if the device is not to be used
for an extended period.
If rechargeable batteries have been stored for a prolonged period, their high discharge rate
may mean that they are discharged or defective (see “5.8 Use rechargeable batteries” on
page 46).
9.2 Disposal
Information
You should also follow the applicable hygiene regulations for sleep laboratories/
clinical facilities.
Dispose of device and USB cable
Do not dispose of the device and the USB cable in domestic waste. To dispose of
the device properly, contact an approved, certified electronics scrap dealer. You can
obtain the address from your Environment Officer or your local authority. The
device packaging (cardboard and inserts) can be disposed of in paper recycling
facilities.