Function check
53
EN
5. Check whether the measured values shown in the display are plausible.
The device is working correctly if the measured values
shown in the display are plausible.
Check display
1. Put on the diagnostic nasal cannula (see “4.2 Put on diagnostic nasal cannula” on
page 31) and/or the pulsoximetry sensor (see “4.3 Attach pulsoximetry sensor” on
page 33).
2. Press the button on the device.
The display is working correctly if the signals from the sensor are shown in the display.
Check sensors
1. Put on the device (see “4.1 Put on device” on page 31).
2. Put on the diagnostic nasal cannula (see “4.2 Put on diagnostic nasal cannula” on
page 31) and/or the pulsoximetry sensor (see “4.3 Attach pulsoximetry sensor” on
page 33).
3. Press the button on the device.
4. Check whether the measured values shown in the display are plausible.
The checked sensor is working correctly if:
• the respiratory flow display in the display is moving.
• the measured values for oxygen saturation and pulse frequency are plausible.