Troubleshooting
55
EN
8.2 Error messages in the display
If a critical fault is detected when the device is switched on, the following error message
appears in the display instead of the start screen:
In the event of one of these faults, return the device to the manufacturer to have it
checked.
No connection can be made to the
device.
Cable connection between
device and PC is interrupted
Check all cable connections.
Device has switched off.
Connection has been inactive for
over 5 minutes
Switch device back on at the
button.
No manual recording can be
started.
The device has been
programmed for a certain
recording time.
Erase data (see “5.5 After the
recording” on page 39). If
necessary, read out the data
present in the device
beforehand.
Error message
Fault
ERROR 1
Program code
ERROR 2
RAM
ERROR 3
UART
ERROR 4
I2C
ERROR 5
SPI
ERROR 6
USB
ERROR 101
Real Time Clock
ERROR 102
EEPROM
ERROR 103
Data Flash
ERROR 104
Temperature sensor
ERROR 105
Display