Alarm settings

NOTE : totalisation stops during alarm adjustment

Alarm 1 and 3  relate to the accumulated total. Alarm 2 and 4 relate to pulsed
outputs.

If you press the ALARMS button momentarily, you can view each of the 4 alarm
settings (each press will illuminate in turn AL1, AL2, AL3 and AL4  LEDs).
alarm settings are not locked out by the lockout switch.

To change alarm settings, select the alarm you  wish to change as shown above
until its LED is flashing, then press the ALARM key for more than 3 seconds.
You will see one digit is brighter than the others. You can change its value using
the UP/DOWN buttons, and then select other digits with the DIGIT pushbutton.
When the value has been set, press OK.

The alarm action is now displayed. This will show ’HI’ for HIGH alarm action,
or  ‘off’ for NO alarm action. You can change this with the UP/DOWN buttons.
Press OK when set.

Alarm channel LED will
flash.

Here, we see that Alarm
1 has its setpoint at
250.00

DIGIT

Min.

ZERO

SPAN / dp

ZERO

SPAN & d.p.

MAX / MIN

TARE

ALARMS

OK

RESET

OUTPUT

AL1

AL2

AL3

AL4

0% O/P                100% O/P

Lineariser

Max.

Содержание INT-AH

Страница 1: ...2003 IEC 1010 89 336 EEC Ampere Time Display Meter Plating Current monitor with pulsed outputs Model INT AH Rev AH 1 1 Software Version DIGIT Min ZERO SPAN dp ZERO SPAN d p MAX MIN TARE ALARMS OK RES...

Страница 2: ...Notes...

Страница 3: ...Calibration Tamper detection 8 Connections 5 Decimal point position selection 6 Declaration of Conformity 17 Failsafe alarm setting 10 General description 3 Getting Started 4 Installing options 9 Intr...

Страница 4: ...cover damage caused by misuse or accident IMPORTANT If the meter is a vital component in your process you may wish to buy a spare to cover possible failure or accident as we cannot guarantee instant...

Страница 5: ...heck that model number and supply voltage suit your application before you install the meter Don t touch any circuitry when the meter is powered because lethal voltages may be present on the circuit b...

Страница 6: ...n when fitted with appropriate plug in option boards Alarm outputs 2 and 4 give pulsed outputs of 100mS each time a batch total has been reached The front panel has a 6 digit 7 segment window for disp...

Страница 7: ...hesignalandpowercables totheappropriatescrewterminalconnectors Check that you are using the correct terminals or you may cause damage to the meter Do not connect any output or alarm cabling yet Apply...

Страница 8: ...unt should be located in the low voltage unswitched return line to minimise common mode noise These connectors are only used when options are fitted Connections E Common No function Disp Current Reset...

Страница 9: ...or ampere hour applications this is normally the rated current of your shunt If the value is correct press the OK button otherwise use the DIGIT key to select digits and the UP or DOWN arrow keys to a...

Страница 10: ...you want to count ampere HOURS b Set InHi 75 00 mV c Set rd Hi to 200 0 d Set InLO to 0 00 mV e Set lo Rd to 0 2 You have a 60mV shunt rated at 100 Amperes and want to display ampere hours to the nea...

Страница 11: ...ommand Front Panel key reset command 1 Link terminal 7 to terminal 10 2 Set lockout switch ON 3 Press Down Arrow key to reset display Remote contact closure resetting 1 Connect a normally open contact...

Страница 12: ...board We strongly suggest that you use anti static precautions toprevent damage to the semiconductors The board assemblies will look something like this Always be careful to connect the pins to socket...

Страница 13: ...ill see two small screws one on each surface Remove both screws Now clip off the front bezel and slide the meter assembly carefully out via the front of the case The relay board plugs into the main bo...

Страница 14: ...e until its LED is flashing then press the ALARM key for more than 3 seconds You will see one digit is brighter than the others You can change its value using the UP DOWN buttons and then select other...

Страница 15: ...sion Dual slope 10 conversions per second Resolution 1 in 63000 max over full range Display update rate 2 5 readings per second Memory of total E2PROM with 10 year storage No batteries required Plug I...

Страница 16: ...Page 13 Record of Revisions Changes 31July2003 Released...

Страница 17: ...Notes Page 14...

Страница 18: ...Page 15 48 0 96 0 92 0 22 0 102 0 12 0 DIGIT Min ZERO SPAN dp ZERO SPAN d p MAX MIN TARE ALARMS OK RESET OUTPUT AL1 AL2 AL3 AL4 0 O P 100 O P Lineariser Max...

Страница 19: ...Page 16 12 0 102 0 22 0 44 0 92 0 45 0...

Страница 20: ...ngshallberoutedseparatelytopowercarryingcabling includesrelayoutputwiring Allsignalcablingshallbescreened Thescreenshallonlybeterminatedtothepowerearthterminal This is to confirm that the Product cove...

Отзывы: