5 Movement limitations
The movements of the operating table are limited by the software. The table has reached a limit, when the
movement stops and a signal can be heard. The limitations are presented on the table below (Table 3). The
first line of the Position/Adjustment column presents the min/max position/adjustment in normal situation.
The min/max values of special situations are presented on the following rows.
Table 3: Movement limitations
Function
Position/Adjustment Situation
Height of the table
65 cm
Normal minimum height
98 cm
Minimum height, when the leg section angle is > 65°
down.
Trendelenburg
-30°
Normal maximum angle
-20°
Maximum angle, when the tilt angle > ±5°.
-20°
Maximum angle, when the back section angle is -20° and
the height is > 80 cm.
-10°
Maximum angle, when the back section angle is -20° and
the height is < 80 cm.
Anti-trendelenburg
-30°
Normal maximum angle
-20°
Maximum angle, when the tilt angle > ±5°.
-17°
Maximum angle, when the leg section angle is > 65°
down and the height is > 80 cm.
-12°
Maximum angle, when the leg section angle is > 50°
down and the height is < 80 cm.
Tilt
±17°
Normal maximum angle
±20°
Maximum tilt in horizontal position.
±5°
Maximum tilt when the trendelenburg/anti-trendelenburg
angle is >20°
Disabled
Leg section angle is > 65° down.
Leg section
-90°
Normal maximum angle down
-65°
Maximum angle when the height of the table is <98 cm.
Slide
30 cm
Normal slide towards the long end of the base frame.
Disabled
Trendelenburg/Anti-trendelenburg >13°
Leg section angle is > 1° down.