6 Cleaning and disinfecting
Before cleaning remove all accessories
Make sure that the charging cable is unplugged.
Clean stains as soon as possible.
In order to keep the surfaces in good condition do the cleaning regularly. Do cleaning/disinfectant
always between patients. Do more thorough cleaning once a month. Follow the cleaning/disinfecting
instructions given by the respective facility.
Metal and plastic surfaces
Clean the metal and plastic surfaces and the hand controls with a damp cloth and weak alkaline cleaning
fluid. Use small brush for corners and other difficult spots. Rinse with clean water and dry carefully after
cleaning. Do not use excessive fluids
Use disinfectant (alcohol or chlorine)
and follow the disinfectant manufacturer’s instructions for use. Let dry
by evaporation in room temperature.
Plastic surfaces (ABS, HDPE, PP)
are highly resistant to chemicals. Plastic is resistant to bleaching agents
(alkaline compounds), dilute organic or inorganic acids. Also solvents and cleaning agents may be used.
Plastic surfaces may get damaged if aromatic hydrocarbons (benzene and its derivates), ketones, ethers,
esters and chlorinated hydrogarbons are used. Plastic might also deteriorate if it is exposed to various
chemicals at the same time.
Stainless steel surfaces
are highly resistant to chemicals. Use for mild detergent solution for cleaning.
Ammonia and most of the solvents can be used to remove difficult stains. Avoid chlorine based solutions.
Painted or chromed metal surfaces
can be cleaned with mild detergent. They are also highly resistant to
chemicals. Do not use harsh abrasive powders on these surfaces.
All surfaces must be dry before connecting to mains and using the device.
Turn the main switch to position OFF.
Always unplug the charging cable before cleaning.
Do not use water spray (shower, high-pressure water guns) for cleaning. Do not use steam or
extremely hot water.
Do not clean in high temperature and air humidity. This might cause interference in electric or
hydraulic system.
Do not expose the device to excessive moisture which can result in liquid pooling.
Do not use solvents or petrol for cleaning. Do not use acids for cleaning.
Always dry the surfaces after cleaning and disinfecting.
Disinfecting wears out the surfaces. After disinfecting clean the surfaces with clean, damp
cloth.
Dilute the disinfectant according to the manufacturer’s instructions.
Do not let any liquids enter the main connection panel or the inner parts of the table.