Configuring the UTRED-16 for logging
UTRED-16 User Guide
14
Figure 2:
UTRED-16 advanced configuration screen in LogTag
®
Analyzer 3
Parameters that influence the appearance of the display are:
l
Pause alarm/statistics processing (see
l
Switch off display after 30 seconds (Power save, see
)
l
allowing a user to reset the trip's minimum and maximum values on the display
during recording
l
showing the alarm screens during review (
Parameters influencing specific behavior of the UTRED-16 are:
l
clearing and resetting alarms when the
START
/Mark
button is pressed
l
leaving the alarm turned on, even if readings return to the normal temperature range
again
l
allowing the user to stop the logger with the
STOP
/Review
button
l
allowing the user to reset the logger with the
START
/Mark
button
For detailed information about each parameter please read the section about
Configuring a
LogTag
®
for logging
in LogTag
®
Analyzer’s User Guide or press F1 for help.
Alarm Configuration Options
LogTag
®
Analyzer allows up to 6 different alarm trigger conditions for temperature when
configuring a UTRED-16 for recording, depending on the model.
Each alarm trigger condition consists of a threshold temperature value, an activation type
(which can be instant, consecutive or accumulative) and a delay time, if it is not an instant
alarm.
If an alarm trigger condition requires readings to exceed an upper threshold temperature it
is called an
upper alarm
. If an alarm trigger condition requires readings to go below lower
thresholds it is called a
lower alarm
.
An alarm event is generated, when either of the entered alarm conditions is triggered.
The UTRED-16 can have up to three upper alarms, and up to three lower alarms.