Operating the Logger
UTRED-16 User Guide
22
During Recording
During normal operation the display shows the most recently recorded temperature. The
current time is also displayed in 24-hour format. A tick symbol is shown as long as no
alarm event has occurred.
Temperatures can be shown in Celsius (as on the left) or Fahrenheit (as on the right).
The temperature unit on the display is the same as used for the PDF and CSV files.
It is set during configuration.
Readings above or below alarm thresholds
If the most recently recorded temperature is above or below one of the alarm thresholds,
one or more alarm markers will be shown on the display.
This symbol is displayed ...
... when the most recent temperature recorded is
above the primary upper alarm threshold, but below the
secondary
above the secondary upper alarm threshold, but below the
tertiary
above the tertiary upper alarm threshold
below the primary lower alarm threshold, but above the
secondary
below the secondary lower alarm threshold, but above the
tertiary
below the tertiary lower alarm threshold
The presence of an alarm marker on the display does NOT indicate an alarm event.
It only shows that the most recent temperature recorded was above or below the
corresponding alarm threshold value. If an alarm marker is shown, but the display
still shows a tick, a delayed alarm is configured.
In this example, the most recently recorded temperature
is above the primary upper alarm threshold, but below a
secondary threshold and no alarm was triggered: