Operating the Logger
UTRED-16 User Guide
33
Stopping the UTRED-16
Automatically
The UTRED-16 automatically stops recording temperature when the maximum number of
readings specified during configuration has been reached. Your unit can also be set up to
stop when it is inserted into a USB socket. This option needs to be set up at the factory
and cannot be changed during configuration with LogTag
®
Analyzer. Your distributor can
supply more information about this option.
Manually
You can configure a UTRED-16 so it can be stopped with the button. This feature is useful
when you take the logger out of a shipment and don’t want to falsify the statistics with
readings taken after the shipment completion. The stop function is enabled in the
Advanced Options
dialogue during configuration.
Press and hold the button. The
STOPPED
symbol shows in
addition to
for 2 seconds.
The
symbol turns off after 2 seconds. Release the
button and the logger stops taking temperature readings.
If the button is released while
is still on, or you wait until
STOPPED
disappears, the
display shows
again, and the logger continues to record data. In this case, an
inspection mark will be recorded in the data against the next temperature reading. The
MARK
symbol will show, and the
PAUSED
symbol also shows if this feature is
.
Unlike many other LogTag
®
products with display, the alarm does
not
have to be
cleared before the logger can be stopped.
Logger has stopped
Once the logger has stopped, the display will show:
l
the
STOPPED
symbol
l
the OK indicator ( ) if no alarm was recorded
l
the alarm indicator ( ) and one or both alarm threshold( ) indicators if an alarm
was recorded during the trip that has not been cleared
l
the current time
l
the battery low symbol ( ) if the battery is low
The logger does not show a temperature reading.