Status (Pop-up window)
Meaning
How to change the status
OptiCare X integrated mattress does not
work because of the bed disconnected
from the mains power.
Connect the bed to the mains
power to enable functions of the
OptiCare X integrated mattress!
At least one air hose is disconnected
from the OptiCare X SCU.
Press MUTE Icon to mute acous-
tic alert sounding from SCU.
Connect all the air hoses to the
OptiCare X SCU.
Integrated Mattress infl ation failed!
Press MUTE Icon to mute acous-
tic alert sounding from SCU. Clo-
se manual CPR valve or check
air connections to SCU to enable
mattress infl ation! If problem is
still present, transfer patient to
another surface!
Integrated Mattress defl ation failed! Auto-
matic defl ation is not available!
Press MUTE Icon to mute acous-
tic alert sounding from SCU.
Open manual CPR valve to ena-
ble mattress defl ation! If problem
is still present, transfer patient to
another surface!
Mattress Identifi cation failed! Connected
mattress is not compatible with OptiCare
X SCU!
Press MUTE Icon to mute acous-
tic alert sounding from SCU. Only
connect the compatible mattress
to the OptiCare X SCU! Check
hoses in mattress near connecti-
on and re-connect the OptiCare X
integrated mattress!
Mattress Identifi cation failed due to a
clogged or kinked air hose!
Press MUTE Icon to mute acous-
tic alert sounding from SCU.
Ensure the air hoses are not
kinked and remove found kinks
and blockages! Check hoses in
mattress near connection and re-
-connect the OptiCare X integra-
ted mattress!
P100
Bed disconnected. Mattress not
operating. Plug bed into socket.
Mattress disconnected.
Check air connectors.
P101
Mattress inflation failure.
Close manual CPR.
P102
P105
Use manual CPR.
Mattress identification failed.
Use compatible mattress only.
P106
Check pipes for kinks or blockages.
P107