4. 

 Press the mode button to confirm Calibration 
within 10 seconds.

Note:

 If the mode button is not pressed within 

10 seconds then the instrument mode will change 

to Gas Test Mode.

5. 

 G-TECTA

®

 SG will display a progress bar.

6. 

 Apply the gas whilst the progress bar is 
counting down.  G-TECTA

®

 SG will display 

‘PASS’ or ‘FAIL’.

Note:

 You can abort the Calibration test by 

pressing the mode button at any time whilst the 

test is in progress.

The G-TECTA

®

 SG instrument will adjust the value on 

the gas sensor to match the stored calibration gas 

value within the sensor module. If the G-TECTA

®

 SG 

unit does not calibrate successfully, G-TECTA

®

 SG 

will display ‘FAIL’. Your G-TECTA

®

 SG unit must be 

re-calibrated or the sensor replaced.

Gas Test ‘pass’ and ‘fail’ and calibration ‘pass’ and 

‘fail’ values are stored in the Event log.

WARNING

One button Calibration will 

check for small drifts of stored calibration value. 

The Linde Group recommends G-TECTA

®

 SG is sent 

for full certified calibration at six monthly intervals.

G-TECTA Ozone Instrument

Due to the reactive nature of ozone (O

3

) special 

procedures should be followed when calibrating 

gas detectors that incorporate an ozone sensor.

Fittings

For ozone calibrations all fittings and pipe work 

should be stainless steel, brass, aluminium or PTFE. 

Fittings or pipe work in other plastic materials such 

as Tygon must not be used.

Gas flow rate

It is important that the flow rate is set correctly as 

otherwise pressure effects may distort calibration 

values and prevent the gas detector from working 

correctly. To calibrate, the flow rate should be set 

between 0.8-1.0 litres / minute (0.03-0.04 cubic 

feet / minute)

Normal safety precautions for handling ozone 

should always be observed along with any special 

instructions that accompany the calibration gas 

cylinder or generator being used.

G-TECTA

®

 SG CO

2

 Instrument*

When G-TECTA

®

 SG CO

2

 is zeroed in fresh air it 

automatically sets the CO

2

 baseline level to 0.04%. 

When zeroing the instrument in air, ensure you are 

outside in clean air well away from a building or 

any CO

2

 emissions. Please ensure the instrument 

is held well away from the breathing zone of  

the operator.

G-TECTA

®

 SG CO

2

 can also be set to zero in nitrogen 

by using G-TECTA

®

 PC software.**

This is a zero point calibration rather than a  

normal zero. In the PC software, set the calibration 

value to 0.00%, apply nitrogen gas and click the 

Calibrate button.

Note: 

After the zero point calibration is completed, 

remember to re-set the calibration gas level to its 

previous level.

* Also read the section – Instructions – specific for use in non-hazardous areas (Page 4) 
** Recommended calibration gas is 2% CO

2

 in a background of nitrogen with a regulated flow of 0.5 l/min (1 SCFH). 

Care should be taken to vent the nitrogen flow away from the operator and not allow nitrogen flow for longer than 
necessary. If indoors, it is recommended to vent gas out of a window or into an extracted fume cupboard.

Содержание G-TECTA SG

Страница 1: ...G TECTA SG Premium Single Gas Detection Instrument Operating Manual UK...

Страница 2: ...tion given in both our printed and online publications are accurate at the time of issue full technical specifications are not necessarily included Furthermore The Linde Group policy is one of continu...

Страница 3: ...call The Linde Group WARNING The accuracy of G TECTA SG instruments equipped with a toxic gas sensor should be checked periodically with known concentration calibration gas Failure to check accuracy...

Страница 4: ...in a non hazardous safe area Users can only change non rechargeable batteries in G TECTA SG instruments 9 The equipment is not certified for use in atmospheres containing more than 21 Oxygen Area Clas...

Страница 5: ...e button and a high quality display with automatic backlight The instrument continuously monitors the gas level in the environment providing current peak and time weighted average TWA readings for the...

Страница 6: ...ill display the warning message CAL When CAL is displayed the instrument can still function as a standard feature although we strongly recommend that the instrument is sent for calibration as soon as...

Страница 7: ...minutes or LTWA 8 hours WARNING TWA alarms cannot be cleared TWA alarms are functional for toxic gas instruments only Display Menu Options The G TECTA SG menu is displayed when you press the mode butt...

Страница 8: ...levels exceed 100 LEL G TECTA SG locks into alarm and displays 9999 showing an over range condition G TECTA SG will temporarily cut off power to the sensor to prevent burn out and display a progress...

Страница 9: ...The Event log can record over 4800 events Batteries Rechargeable Batteries The re chargeable batteries in the G TECTA SG are Li Ion batteries The batteries will re charge in less than 6 hours when ful...

Страница 10: ...attery pack 1 Ensure you are in a safe non hazardous area 2 Always switch the G TECTA SG off before opening the case to change the battery 3 Remove the back cover and remove the battery pack 4 Replace...

Страница 11: ...sample to reach the sensor Continue to squeeze the bulb for an additional 45 seconds or until gas readings stabilise 6 To remove the flow cap gently unclip the cap from the sensor and lift it away fr...

Страница 12: ...ng down G TECTA SG will display PASS or FAIL Note In the event G TECTA SG displays FAIL please see the bump test calibration troubleshooting guide or contact The Linde Group 6 To abort the gas test pr...

Страница 13: ...as flow rate It is important that the flow rate is set correctly as otherwise pressure effects may distort calibration values and prevent the gas detector from working correctly To calibrate the flow...

Страница 14: ...instrument Gas flow not started in time Repeat gas test Start gas immediately Instrument fails calibration Gas cylinder empty Check cylinder gauge Replace cylinder Gas cylinder out of date Check cylin...

Страница 15: ...CTA PC software options screen and note the ports available ii Switch on the G TECTA SG instrument and drop it into the charger cradle iii Check the available communications ports available again and...

Страница 16: ...ds Visible alarms Dual colour Blue Red flashing LEDs Vibrating alarm Internal vibrating Repeatability 2 FSD 6 months Electromagnetic Compatibility 2004 108 EC Explosion Protection Certification IECEx...

Страница 17: ...ralian ASSG1061 G TECTA SG 230v Battery Charger UK ASSG1062 G TECTA SG 230v Battery Charger Europe ASSG1063 G TECTA SG 110v Battery Charger USA ASSG1064 G TECTA SG 230v Battery Charger Australian c w...

Страница 18: ...ailure Do NOT use instrument Exit hazardous area immediately and return instrument for service Auto Zero failure Instrument zeroed in contaminated atmosphere Switch off instrument and restart in clean...

Страница 19: ...es or oxygen contain chemicals Extreme levels of humidity can cause problems The sensors are rated for an average ambient humidity of 15 90 R H However the sensors are commonly used from the tropics t...

Страница 20: ...seefa 1180 Buxton UK Provisions of the Directive fulfilled by the products Group II Category 2G Ex ia d IIC T4 Gb 20 C Ta 65 C IP65 Flammable Group II Category 1G Ex ia IIC T4 Ga 20 C Ta 65 C IP65 Oxy...

Страница 21: ...Y OF FITNESS FOR A PARTICULAR PURPOSE The Linde Group WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY...

Страница 22: ...22 UK Notes...

Страница 23: ...23 UK Notes...

Страница 24: ...f going to print Denmark AGA A S Vermlandsgade 55 2300 K benhavn S Tel 45 32 83 66 00 Estonia Eesti AGA AS P rnu mnt 141 11314 Tallinn Tel 372 6504 500 Finland Oy AGA Ab Itsehallintokuja 6 02600 Espoo...

Отзывы: