The Linde Group manufactures and supplies  

Gas Test Cylinders under numerous global product 

brands, including Linde, BOC, AGA, Afrox etc. 

The Linde Group gases supplied for use with your 

G-TECTA

®

 gas detection instrument are specially 

formulated, high stability long life single gas mixes 

The G-TECTA

®

 SG instrument will determine 

Pass/Fail status for the gas test.

In order to perform successful gas tests ensure:

•  The gas used has the recommended gas 

concentration and it is within the validity date 

specified by The Linde Group

•  The gas flow path is leak tight. It is important 

to check that the flow cap is properly fitted to 

the G-TECTA

®

 SG unit, the outlet tubing is not 

restricted in any way, nor additional tubing 

length used

The G-TECTA

®

 Gas Test Kit comprises a full gas 

cylinder, a regulator, interconnecting tubing, a 

gas test key (used to activate Gas Test Mode 

on the instrument), a flow cap to attach to the 

G-TECTA

®

 SG, and a vent line. 

The Gas Test Kit is supplied in a convenient  

carry case.

How to Perform a Bump Test

1. 

 Ensure the G-TECTA

®

 SG instrument is 

switched on and in normal operating mode.

2. 

 Fit the flow cap onto the front of the sensor 
and attach the hose from the regulator. Attach 
the outlet hose to ‘vent gas away’ – do not 
extend this hose, restrict or allow kinks.

3. 

 Place the Gas Test Key, vertically aligned, 
onto the instrument, to the left side of the 
display. The G-TECTA

®

 SG will activate the Gas 

Test Mode and show ‘TEST’ on the display.

4. 

 G-TECTA

®

 SG will display a progress bar.

5. 

 Apply the gas whilst the progress bar is 
counting down. G-TECTA

®

 SG will display 

PASS or FAIL.

Note:

 In the event G-TECTA

®

 SG displays 

FAIL, please see the bump test/calibration 
troubleshooting guide or contact The Linde Group.
6. 

 To abort the gas test press the mode button 
at any time whilst the test is in progress.

Instrument Calibration

Calibration of the gas sensor should be performed 

regularly every 6 months. Over time the accuracy 

of the gas sensor can drift outside of what is 

considered a safe range and regular calibration is 

essential to ensure accuracy of gas readings.

There are three methods of instrument calibration 

available. G-TECTA

®

 SG calibration can be 

performed directly on each instrument (one button 

calibration), by utilising G-TECTA

®

 SG PC software 

or by using the Linde G-DOCKA

®

 Hub. Certified 

calibration gases should be used.

WARNING

The accuracy of G-TECTA

®

 SG 

instruments should be checked periodically with 

known-concentration of calibration gas. Failure 

to check accuracy can lead to inaccurate and 

potentially dangerous readings.

How to Perform a One Button Calibration Test 

1. 

 Ensure you are in clean air and the 
instrument is in normal operating mode. 
Double click the mode button and select 
Zero from the options menu by double 
clicking on the Zero icon. G-TECTA

®

 SG will 

perform an Auto Zero. 

To perform a one button Calibration, 
complete the next steps within 15 
minutes of completing the Auto Zero.

2. 

 Fit the flow cap onto the front of the sensor 
and attach the hose from the regulator. Attach 
the outlet hose to ‘vent gas away’ – do not 
extend this hose, restrict or allow kinks.

3. 

 Place the Gas Test Key, vertically aligned, 
onto the instrument, to the left side of the 
display. The G-TECTA

®

 SG will activate the 

Calibration Mode and show “CAL” and “????” 
alternatively on the display.

Содержание G-TECTA SG

Страница 1: ...G TECTA SG Premium Single Gas Detection Instrument Operating Manual UK...

Страница 2: ...tion given in both our printed and online publications are accurate at the time of issue full technical specifications are not necessarily included Furthermore The Linde Group policy is one of continu...

Страница 3: ...call The Linde Group WARNING The accuracy of G TECTA SG instruments equipped with a toxic gas sensor should be checked periodically with known concentration calibration gas Failure to check accuracy...

Страница 4: ...in a non hazardous safe area Users can only change non rechargeable batteries in G TECTA SG instruments 9 The equipment is not certified for use in atmospheres containing more than 21 Oxygen Area Clas...

Страница 5: ...e button and a high quality display with automatic backlight The instrument continuously monitors the gas level in the environment providing current peak and time weighted average TWA readings for the...

Страница 6: ...ill display the warning message CAL When CAL is displayed the instrument can still function as a standard feature although we strongly recommend that the instrument is sent for calibration as soon as...

Страница 7: ...minutes or LTWA 8 hours WARNING TWA alarms cannot be cleared TWA alarms are functional for toxic gas instruments only Display Menu Options The G TECTA SG menu is displayed when you press the mode butt...

Страница 8: ...levels exceed 100 LEL G TECTA SG locks into alarm and displays 9999 showing an over range condition G TECTA SG will temporarily cut off power to the sensor to prevent burn out and display a progress...

Страница 9: ...The Event log can record over 4800 events Batteries Rechargeable Batteries The re chargeable batteries in the G TECTA SG are Li Ion batteries The batteries will re charge in less than 6 hours when ful...

Страница 10: ...attery pack 1 Ensure you are in a safe non hazardous area 2 Always switch the G TECTA SG off before opening the case to change the battery 3 Remove the back cover and remove the battery pack 4 Replace...

Страница 11: ...sample to reach the sensor Continue to squeeze the bulb for an additional 45 seconds or until gas readings stabilise 6 To remove the flow cap gently unclip the cap from the sensor and lift it away fr...

Страница 12: ...ng down G TECTA SG will display PASS or FAIL Note In the event G TECTA SG displays FAIL please see the bump test calibration troubleshooting guide or contact The Linde Group 6 To abort the gas test pr...

Страница 13: ...as flow rate It is important that the flow rate is set correctly as otherwise pressure effects may distort calibration values and prevent the gas detector from working correctly To calibrate the flow...

Страница 14: ...instrument Gas flow not started in time Repeat gas test Start gas immediately Instrument fails calibration Gas cylinder empty Check cylinder gauge Replace cylinder Gas cylinder out of date Check cylin...

Страница 15: ...CTA PC software options screen and note the ports available ii Switch on the G TECTA SG instrument and drop it into the charger cradle iii Check the available communications ports available again and...

Страница 16: ...ds Visible alarms Dual colour Blue Red flashing LEDs Vibrating alarm Internal vibrating Repeatability 2 FSD 6 months Electromagnetic Compatibility 2004 108 EC Explosion Protection Certification IECEx...

Страница 17: ...ralian ASSG1061 G TECTA SG 230v Battery Charger UK ASSG1062 G TECTA SG 230v Battery Charger Europe ASSG1063 G TECTA SG 110v Battery Charger USA ASSG1064 G TECTA SG 230v Battery Charger Australian c w...

Страница 18: ...ailure Do NOT use instrument Exit hazardous area immediately and return instrument for service Auto Zero failure Instrument zeroed in contaminated atmosphere Switch off instrument and restart in clean...

Страница 19: ...es or oxygen contain chemicals Extreme levels of humidity can cause problems The sensors are rated for an average ambient humidity of 15 90 R H However the sensors are commonly used from the tropics t...

Страница 20: ...seefa 1180 Buxton UK Provisions of the Directive fulfilled by the products Group II Category 2G Ex ia d IIC T4 Gb 20 C Ta 65 C IP65 Flammable Group II Category 1G Ex ia IIC T4 Ga 20 C Ta 65 C IP65 Oxy...

Страница 21: ...Y OF FITNESS FOR A PARTICULAR PURPOSE The Linde Group WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY...

Страница 22: ...22 UK Notes...

Страница 23: ...23 UK Notes...

Страница 24: ...f going to print Denmark AGA A S Vermlandsgade 55 2300 K benhavn S Tel 45 32 83 66 00 Estonia Eesti AGA AS P rnu mnt 141 11314 Tallinn Tel 372 6504 500 Finland Oy AGA Ab Itsehallintokuja 6 02600 Espoo...

Отзывы: