Messages
ICON
MESSAGE
DESCRIPTION/ACTION REQUIRED
-
The chamber door is locked; no action required.
The water level inside the clean water tank is below the minimum. Fill the clean water tank.
The water level inside the used water tank is at maximum level. Drain used water tank.
Check the external sterilizer water supply. You might have to replace filter cartridges – drain clean water tank and
follow instructions for use of water filtration system.
The distilled/demineralized water in the clean water tank is of poor quality. Drain the tank and refill it with water of
good quality; refer to ANNEX 7
Don’t touch the chamber or the load with bare hands: high temperature, risk of burns!
The bacteriological filter needs to be replaced.
The dust filter needs to be replaced.
The door seal needs to be replaced.
The 4000 cycle overhaul needs to be performed. Call for service.
The CPU board battery needs to be replaced. Call for service.
PC/Logger not detected (disconnected or not powered).
Cycle report printer configured but not detected (disconnected or not powered).
Label printer configured but not detected (disconnected or not powered).
File saving error (check presence and connection of the USB drive).
Error in the USB communication (between the main board and the USB4 board)
Error in savng the HTML cycle report
Error in savng the SCL cycle report
Bad USB device
NOTE: for any message not listed in this table, call service.