Controls and commands
14
Switching ON the sterilizer
Press the mains switch behind the service door to switch ON the sterilizer.
The visual indicator on the mains switch turns green and the START screen (see next page) appears.
“SLEEP” mode
In “SLEEP” mode the display remains dark and the sterilizer chamber is no longer heated to save energy.
If the sterilizer is not used for 12 hours, (the time interval can be changed, see Chapter 5 - Programming) it will automatically switch to
“SLEEP” mode.
Exit from “SLEEP” mode through any of the following actions:
- Press any button on the control panel;
- Open or (if it is open) close the chamber door.
It is also possible to put the sterilizer into “SLEEP” mode manually:
On the START screen, press the BACK button.
A 10 second countdown will commence. At the end of the countdown the sterilizer will enter the
“SLEEP” mode.
The countdown can be stopped at any time by pressing the BACK button.