Care and Maintenance
For safe and trouble-free operation, a few routine procedures should be performed every day the lift is used.
• Visually inspect the patient lift and check for external damage or wear.
Visually inspect the patient lift and check for external damage or wear.
• Check that all screws and lock nuts on the slingbar are tight.
Check that all screws and lock nuts on the slingbar are tight.
• Check that the safety latches on the slingbar work properly.
Check that the safety latches on the slingbar work properly.
• Check the handcontrol and the lift and width movements.
Check the handcontrol and the lift and width movements.
• Check the emergency lowering function (both electrical and mechanical).
Check the emergency lowering function (both electrical and mechanical).
• Charge the battery every day the lift is used, and check charger function.
Charge the battery every day the lift is used, and check charger function.
When necessary, clean the lift with common surface cleaners or disinfectants and check to ensure that the wheels
are free of dirt and hair.
NOTE! Avoid regular use of cleaning agents that contain phenol or chlorine, since
these can damage aluminium and polyamide material.
Transportation and storage
During transportation, or when the patient lift is not to be used for some time, the emergency stop button should be
pushed in. Store the lift at a temperature exceeding freezing point and at normal relative humidity (no more than
60%).
Inspection
Uno 100/102 must be inspected at least once per year. Pay particular attention to parts that are subject to wear.
Service must be carried out according to Liko service manual and by authorized personnel.
IMPORTANT!
Repairs and maintenance may only be carried out by authorized Liko service personnel using original Liko spare
parts.
Service agreements
Liko invites you to sign a service agreement for regular maintenance and testing of your Uno mobile lift.
Liko's products are constantly being updated and refined. Liko reserves the right to change aspects of the products
without prior notice. Contact your local Liko representative for updated information and advice.
Recycling
For instructions on how the lift should be recycled, please visit our website www.liko.com.
Made in Sweden
Liko is quality certified according to ISO 9001 and its equivalence for the medical device industry, ISO 13485.
Liko is also certified according to environmental standard ISO 14001.
© Copyright Liko
AB 2007-05
Manufacturer:
Liko AB
SE-975 92 Luleå
Sweden
Liko Distributor in UK:
Liko (UK) Ltd.
Brunel Centre,
Stroudwater Business Park,
Stonehouse, Gloucester
GL10 3RU, United Kingdom
Phone: +44 (0) 1453 827272
Fax: +44 (0) 1453 828844
Liko Distributor in USA and Canada:
Liko North America
122 Grove Street
Franklin, Ma 02038, USA
Phone: +1 508-553-3993, +1 888-545-6671
Fax: +1 508-528-6642
www.liko.com