7
www.liko.com
Uno 100/102 • 7EN150104-01
Locking the wheels
The rear wheels can be locked for rotation and lateral move
-
ment. To lock the wheels press the brake lever above the
wheel down with your foot. Release the wheels by pressing
the top side of the lever.
During lifting, wheels should remain unlocked so that the lift
may shift to the patient's center of gravity. The wheels should
however be locked if there is a risk for the lift moving into the
patient, for example when lifting from the floor.
User´s Guide
EM
EE
ES
Mechanical Emergency Lowering
Mechanical emergency lowering is performed by turning the
red emergency-lowering cylinder in the direction of the arrows.
Operation
Operate the lift using the pushbuttons on the handcontrol.
For raising and lowering the lift arm:
press and , respectively.
Directional arrows show the direction of movement (up/down).
The thicker arrows on the handcontrol correspond to the maxi-
mal speed and the thinner arrows correspond to a slower
lifting speed (only mod. ES). Lifting motions stops as soon as
the push-button is released.
For base width adjustment (only mod. EE/ES):
Press either of the two push-buttons.
Wider
Narrower
Base width adjustment (only mod. EM)
(only mod. EM)
Width adjustment of the base is done manually with a lever
that locks in a range of positions. Simply release the lever at
the desired width and it locks in place.
To use the Emergency Stop:
Push the red Emergency Stop button on the control box.
To reset the Emergency Stop:
Turn the button in the direction of the arrows.
Electrical emergency lowering
With a narrow object, press the hole marked "EMERGENCY"
on the control box.
Model ES is also equipped for electrical emergency raising.
The object used to press must not be sharp, since
this could cause damage on the control box.
Never pull the lift by the actuator! This may result in
damage to the lift and possible injury.
Locked wheels during lifting increases the risk of the
lift tilting over.
ON
CHARG
EMERGENCY
BATTERY
ON
CHARGE
EMERGENCY
BATTERY
ON
CHARGE
EMERGENCY
BATTERY