Hand control
Indicator lamp (A) - charge the battery of the lift!
The lifting motion is operated with the push buttons on the hand
control. The direction in which the arrows are pointing applies
when the hand control is held as shown in the picture. To raise or
lower the lift arm, press or . For lower lifting speed use the
thinner arrows. The lifting motion stops as soon as the push button
is released. For adjustment of the base width, press or .
Operation
Never move the lift by pulling on the
actuator!
ON CHARG
EMERGENCY
BATTERY
ON CHARGE
EMERGENCY
BATTERY
ON CHARGE
EMERGENCY
BATTERY
Emergency stop
Activate
:
Press the red button on the control box.
Reset
:
Turn the button in the direction shown by
the arrows until the button springs out.
Locking the wheels
The rear wheels can be locked to prevent rotating and turning. To lock
the wheels, push down the lock pedal with your foot. To unlock the
wheels, push on the raised button at the wheel. During passive/active
lifting, the wheels should be unlocked so that the lift can be moved to
the patient’s centre of gravity.
Electrical emergency lowering/raising
Use a narrow object to push the buttons
inside the labelled holes on the control box.
The object used to press must not be
sharp, since this may cause damage on the
control box!
Mechanical emergency lowering
Pull the red emergency lowering
control straight upwards. Mechanical
emergency lowering only works when
the lift arm is under load, i.e. when a
patient is standing/sitting in the lift.
The lowering motion is slightly delayed.
A
Installation of latches
After installation, check that the
latch locks and moves freely in the
sling bar hook.