18
Sabina
™
II sit-to-stand lift • 7EN155106 Rev 1
www.hill-rom.com I www.liko.com
Disinfection Instructions
1. For the use of suitable disinfectants see ”Application of commonly used Cleaning / Disifectants on Liko products” in this
document.
2. Use the choice of disinfectant according to the manufacturer’s instructions and repeat the work step as in”Cleaning
Instructions"
3. Remove traces of disinfectant after disinfection. Wipe off the lift with a cloth moistened with clean water starting from
the top and working down. The cloth shall not be so damp that it drips.
The lift may not be cleaned with CSI or equivalent.
The hand control may not be cleaned with Viraguard or equivalent.
The control box may not be cleaned with Anioxy Spray or equivalent.
The foot rest may not be cleaned with Terralin Protect, Virex II or equivalent.
The leg support on Sabina II, S/N up to 460899 may only be cleaned with warm water and a facility approved
neutral detergent.
The leg support on Sabina II S/N from 460900 may be cleaned with the recommended disinfectants.