Care and Maintenance
For safe and trouble-free operation, a few routine procedures should be performed every day the lift is used:
• Visually inspect the lift and check for external damage or wear
• Check the slingbar connector
• Check the lift strap for wear and make sure it is not twisted
• Check that the safety latches work properly
• Test the operation of raising and lowering and the base width function
• Check that the correct lifting range has been set and that the emergency lowering functions
• Charge the battery every day the lift is being used, and check the charger function
If needed, clean the patient lift with common water-based surface disinfectants
.
Warning! Do not routinely use
chemicals containing phenol or chlorine, since these may damage aluminium and polyamide materials.
Transportation and storage
During transportation, or when the patient lift is not to be used for some time, the emergency button should be
pushed in. Also, when storing, keep the patient lift at a temperature exceeding freezing point and at a relative
humidity no more than 60%.
Inspection
Golvo must be inspected at least once per year. Pay particular attention to parts that are subject to wear.
Inspection must be performed according to the Liko service manual and by authorized personnel.
IMPORTANT!
Repairs and maintenance may only be carried out by personnel authorized by Liko using original Liko spare parts.
Service agreements
Liko invites you to sign a service agreement for regular maintenance and testing of your Golvo mobile lift.
Contact your local Liko representative for more information.
Liko’s products are constantly being updated and refined. Liko reserves the right to change aspects of the
products without prior notice. Contact your local Liko representative for updated information and advice.
Recycling
For instructions on how the Golvo mobile lift should be recycled, please visit our website www.liko.com.
Made in Sweden
Liko is quality certified according to ISO 9001 and its equivalence for the medical device industry, ISO 13485.
Liko is also certified according to environmental standard ISO 14001.
www.liko.com
Manufacturer
Liko AB
Nedre Vägen 100
SE-975 92 Luleå
Sweden
[email protected]
National Distributor in USA and Canada:
Liko Distributor in UK
:
Liko (UK) Ltd.
Brunel Centre,
Stroudwater Business Park
Stonehouse, Gloucester
GL10 3RU, United Kingdom
Phone: +44 (0) 1453 827272
Fax: +44 (0) 1453 828844
[email protected]
Liko Inc
122 Grove Street
Franklin, Ma 02038, USA
Phone: 508-553-3993, 1-888-545-6671
Fax: 508-528-6642
[email protected]