SMA1
Serial number
[44 … 47, ro]
These registers show the serial number of the device expressed in hexadecimal
notation.
Register 44
: year of production
Register 45
: week of production
Registers 46 and 47
: serial number in ascending order
Command
[48, wo]
Value
Function
00
Normal operation
01
Save parameters on EEPROM
02
Save and activate Preset / Offset
04
Load and save default parameters
After having set a new value in some register, use the
function in this register to store it. Set “01” in the register.
After having set a Preset / Offset value, use the
function in this register to both store and activate the preset / offset at
the same time. Set “02” in the register.
Load and save default parameters
: default parameters are set at the factory
by Lika Electronic engineers to allow the operator to run the device for standard
operation in a safe mode. As soon as the command is sent the default
parameters are uploaded and activated. All parameters which have been set
previously are overwritten, thus previously set values are lost. The complete list
of machine data and the relevant default parameters preset by Lika Electronic
engineers are available on page 46. Set “04” in the register.
WARNING
Load and save default parameters
parameters which have been set previously are overwritten, thus previously set
values are lost!
As soon as the command is sent, the register is set back to "00" (
) automatically.
Wait min. 30 ms (EPROM writing time) before using a new function.
Default = 00 (
MAN SMA1 E 2.9.odt
6 - BiSS B-mode interface
24 of 48