Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 – Operators Manual
Rev. No 01
Page 87 of 126
7.2 Double clicks the LIO for entering to realign process
7.3 Enter password “123” and return the ‘Enter’ key
7.4 Initial the “Start” button and realign the X or Y axis accordingly
7.5 Repeat the process until it is centered and save the value before exit it
and validate again by turning off the console and on
Double clicks
the LIO Icon
Initial realignment
process
Initial ‘Save’
Y-axis alignment
X-axis alignment