Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 – Operators Manual
Rev. No 01
Page 40 of 126
1. STANDBY Mode Button
When this indicator is illuminated, the laser system is in the Standby mode and
both of the aiming beam and the laser power can't be delivered to the patient.
2.
READY Mode Button
When this indicator is illuminated, the Laser System is in the READY mode and
both of the aiming beam and the laser power can be delivered to the patient
once the footswitch is pressed or activated. The system will return back to the
STANDBY mode, if the unit is being in non-operation status for 10 minutes or
deactivated. This is also used to provide a ready indicator prior to emissions of
the laser treatment.
3. Power Control Button
This control is used to adjust the laser output power. By default setting the
laser power is at set at zero unless the reconfiguration is made. The setting can
be adjustable from 50 mW to 2000 mW and by pressing the up / down buttons
will increase or decrease the power value. The setting will be reset it when
power is down.
4. Set Power Display
This display will show the set power level and the display will change as the
power is adjusted.
5. Exposure Duration setting switches
These two switches control the duration of the exposure time that the treatment
laser is given to the patient each time the footswitch is pressed in the READY
mode. The left hand control will decrease the duration and the right hand
control will increase the duration. The minimum duration is 0.01 seconds and the
maximum duration is 3.0 seconds. If the duration is adjusted beyond 3.0 seconds,
then the system will enter the continuous pulse mode. In this mode, the laser
will output continuously until the doctor/user releases the footswitch. When this
mode is activated, the Duration display will show “CW”.
6.
Exposure Duration Display
This display shows the exposure duration setting in seconds.
7. Repeat Interval setting switches
These switches are used to set the repeat interval duration, which is the time
between each pulse of the treatment laser. Similar to the Duration switches, the
left hand control will decrease the interval and the right hand control will
Содержание LightLas 532/670
Страница 1: ...Operator Manual Multi Wavelength Pattern Scanning System LightLas 532 670 ...
Страница 4: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Operators Manual Rev No 01 Page 3 of 126 ...
Страница 108: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Operators Manual Rev No 01 Page 107 of 126 ...
Страница 109: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Operators Manual Rev No 01 Page 108 of 126 ...
Страница 110: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Operators Manual Rev No 01 Page 109 of 126 ...
Страница 111: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Operators Manual Rev No 01 Page 110 of 126 ...
Страница 127: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Operators Manual Rev No 01 Page 126 of 126 ...