Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 – Operators Manual
Rev. No 01
Page 112 of 126
B.
Remote Control Warning Only
Warning
Cause(s)
Remedy Action(s)
inL rE-F
-Fiber not connected
-Interlock plug not
connected
-Check the fiber connection
-Check the Interlock connection
-If the problem persists, call the
authorized service or
manufacturer for further
assistance
inL con
-Footswitch not
connected
-Check if the footswitch is
connected
-Check footswitch integrity
-If the problem persists, call the
authorized service or
manufacturer for further
assistance
inL dEL
-Delivery key not
connected
-Check if the delivery key is
plugged in all the way
-If the problem persists, call the
authorized service or
manufacturer for further
assistance