9
300544655_002_C1 - 11/2016 - © Leybold
Mechanical setpoint relays:
P/N: 230350V02, 230351V02, 230354V02, 230355V02:
0
P/N: 230352V02, 230353V02
,
230356V02:
2
P/N: 230366V02:
3
Setpoint relay range:
2×10
-4
to 1300 mbar
Setpoint relay contact rating:
1 A / 30 VDC/AC (resistive load)
Setpoint relay response time:
< 100 ms
Setpoint relay contact resistance:
100 mΩ (max)
Setpoint relay contact endurance:
100,000 cycles (min) (30 VDC/1 A load)
Setpoint relay contact endurance:
2,000,000 cycles (min) (30 VDC/0.2 A load)
Software setpoint relays:
P/N: 230702V02:
2
Response time:
< 20 ms
Digital interface cycle time:
1 ms
Materials exposed to vacuum
(2)
:
304 stainless steel, sealing material FPM
Internal volume:
KF16 flange
2.8 cm
3
NPT 1/8”
3.1 cm
3
Housing material:
Stainless steel 304
Flange material:
Stainless steel 304
Weight:
P/N: 230350V02, 230352V02, 230366V02:
168 g
P/N: 230351V02, 230353V023:
183 g
P/N: 230354V02, 230356V02:
292 g
P/N: 230355V02:
307 g
P/N: 230702V02:
338 g
Maximum allowed pressure:
3 bar
Operating temperature:
0 to 40 °C (32 to 104 °F)
Bake out temperature (Power off):
85 °C (185 °F)
Filament temperature:
35 °C above ambient temperature
Humidity:
0
– 95% Non-condensing
Ingress Protection Rating:
IP40
Leak rate:
< 5
∙
10
-9
mbar
∙
l/s
(1) Accuracy and repeatability are typical values measured in Nitrogen atmosphere after zero adjustment at
ambient temperature.
(2) For the full list of all materials exposed to process gases please contact LEYBOLD.
Содержание THERMOVAC TTR 101 N
Страница 10: ...10 300544655_002_C1 11 2016 Leybold 2 2 Dimensions mm P N 230702V02 P N 230354V02 to 230356V2...
Страница 11: ...11 300544655_002_C1 11 2016 Leybold P N 230366V02 P N 230350V02 to 230353V02 Flanges KF 16 NPT 1 8...
Страница 39: ...39 300544655_002_C1 11 2016 Leybold...
Страница 40: ...40 300544655_002 10 2015 Leybold Declaration of Conformity...
Страница 41: ...41 300544655_002 10 2015 Leybold Notes...
Страница 42: ...42 300544655_002 10 2015 Leybold Sales and Service...