All steps from Normal operation to the different adjustment modes and related adjusting steps can also be found in
the following guide:
Adjustment User switch and LED-ring/LED action
Zero
adjustment
Press button for 1 second until light ring is off
Release button to enter Zero adjustment mode
Light ring is pulsing red and green
Press button >1 second to perform Zero adjustment
Press button <1 second to
go back to Normal
Operation
LED-ring/LED is off
Release button
LED-ring/LED flashes green 10
times:
Zero adjustment succeeded
LED-ring/LED flashes red 10
times:
Zero adjustment failed
Setpoint 1
adjustment
Press button for 1 second until light ring is off
Continue pressing button for 3 seconds until light ring shows single green flash
Release button to enter Setpoint 1 adjustment
LED-ring/LED is pulsing green
Press button for >1 second
Press button <1
second to go back
to Normal
Operation
LED-ring/LED is off
Output current setpoint value (*)
Continue pressing button for >5 seconds
Release button to
keep current voltage
output threshold to
Analog output and
return to Normal
operation
Ramp up/down voltage/setpoint value (*)
Release button to keep current voltage output
threshold to Analog output
Tap button to toggle
through Output voltage
with small increments of
10 mV
Wait for >5 seconds
LED-ring/LED flashes
green
Wait for >5 seconds
LED-ring/LED flashes green 3 times
Selected Output voltage is stored
Setpoint 2
adjustment
Press button for 1 second until light ring is off
Continue pressing button for 6 seconds until light ring shows single red flash
Release button to enter Setpoint 2 adjustment
LED-ring/LED is pulsing red
Press button for >1 second
Press button <1
second to go back to
Normal Operation
LED-ring/LED is off
Output current setpoint value (*)
Continue pressing button for >5 seconds
Release button to
keep current voltage
output threshold to
Analog output and
return to Normal
operation
Ramp up/down voltage/setpoint value (*)
Release button to keep current voltage output
threshold to Analog output
Tap button to toggle
through Output voltage
with small increments
of 10 mV
Wait for >5 seconds
LED-ring/LED flashes
green
Wait for >5 seconds
LED-ring/LED flashes green 3 times
Selected Output voltage is stored
(*) For transmitter versions with integrated touch display (P/N: 230354V02, 230355V02, 230356V02), the pressure
setpoint value is shown on the display.
Содержание THERMOVAC TTR 101 N
Страница 10: ...10 300544655_002_C1 11 2016 Leybold 2 2 Dimensions mm P N 230702V02 P N 230354V02 to 230356V2...
Страница 11: ...11 300544655_002_C1 11 2016 Leybold P N 230366V02 P N 230350V02 to 230353V02 Flanges KF 16 NPT 1 8...
Страница 39: ...39 300544655_002_C1 11 2016 Leybold...
Страница 40: ...40 300544655_002 10 2015 Leybold Declaration of Conformity...
Страница 41: ...41 300544655_002 10 2015 Leybold Notes...
Страница 42: ...42 300544655_002 10 2015 Leybold Sales and Service...