Installation
36
130001398_002_C0 - 09/2016 - © Leybold
3.1.3 Filling in of the Lubricants
Not required for the WH/WHU 2500.
The lubricant needed for running the pump is supplied in a separate contain-
er.
Unscrew the oil-fill plugs and add lubricant. Please note that both oil reser-
voirs (gear and motor side) have to be filled separately. There is no connec-
tion between them.
An oil without additives and of viscosity class ISO VG 100 (formerly SAE 30)
must be used for the pump. We recommend the use of our special oil LVO
210. As PFPE we recommend LVO 400 or LVO 410 sold by us. Please con-
sult us if you intend to run the pump with other oils or special lubricants.
It needs to be ensured that the oil filling levels stated in fig. 3.6 which apply
to switched off pumps (at standstill) are correctly maintained.
NOTICE
RUVAC WH/WHU 4400/7000
Fig. 3.4 Connections and controls on the RUVAC WH 4400/7000
1
4
6
8
5
7
3
2
10
3
7
5
6
4
9
1 Discharge flange
2 Intake flange
3 Oil inlet ports
4 Oil outlet ports (vertical operation)
5 Oil outlet ports (horizontal operation)
6 Oil sight glasses (vertical operation)
7 Oil sight glasses (horizontal operation)
8 Cooling water intake port
9 Cooling water outlet port
10 Direction of rotation arrow
M16 x 1,5
Содержание RUVAC WH 2500