Installation
GA09420_002_C0 – (2016-10) –
Leybold
15
Slide the electronics unit in to the mechanical stop and lock it with the hexa-
gon socket set screw (1).
The electronics unit is now installed.
3.1.3 Using the Optional Baffle
In severely contaminating processes and to protect measurement electrodes opti-
cally against light and fast charged particles, replacement of the built-in grid by the
optional baffle (
→
43) is recommended.
Installing/deinstalling
the baffle
The optional baffle will be installed at the sensor opening of the deinstalled trans-
mitter (Deinstallation
→
36).
Caution
Caution: dirt sensitive area
Touching the product or parts thereof with bare hands increases the
desorption rate.
Always wear clean, lint-free gloves and use clean tools when working
in this area.
Required tools / material
•
Baffle
(
→
43)
•
Pointed
tweezers
•
Pin (e.g. pencil)
•
Screwdriver No 1
Procedure
Carefully remove the grid with tweezers.