Head
GA10218_0101 - 5/2002
⏐
17
Blank flange DN 25 KF
Clamping ring DN 25 KF
Centering ring DN 25 KF
1.4.2 Accessories and Options
The following parts can be ordered additionally:
1.4.2.1 Sniffer line SL300
By use of the sniffer line the PhoeniXL can easily be converted to a sniffer
leak detector. The length of the sniffer line is 4m (i.e. 12 feet).
Installation:
The sniffer line is to be adapted to the KF 25 of the PhoeniXL with the
small flange. Then the plug of the sniffer line is to be connected to the
input „Options“ of the PhoeniXL.
Function:
The green LED is on when
– the PhoeniXL is ready for use and
– the selected trigger level is not exceeded
The red LED is on when
– the PhoeniXL is not ready for use or
– the selected trigger level is exceeded.
The pushbutton in the grip is for the zero function. When pushing the but-
ton the helium background will be suppressed. For cancelling the zero
function push the button once more. (refer to Chapter
4.1.2.4
)
Sniffer Line SL300
252003
Leak Ware (Software for data aquisation)
14090
Helium Sniffer QUICK-TEST QT100
15594
Remote Control Set consisting of
– Remote control
– Cable , 4 m
– Mounting parts
Extension Cable for remote control, 8 m
252002
14022
Spray gun with hose
16555
Set of connection plugs
20028782
Partial flow system PhoeniXL
300
only
14020 for EU-ver-
sion
Partial flow system
14028 for US-ver-
sion
Содержание 250000