9 Fault finding
Table 14 Fault finding
Condition
on page 52
Pump does not reach the ultimate pressure
on page 53
After switching off the pump under vacuum, pressure in the system rises
on page 53
Oil in the intake line or the vacuum vessel
on page 54
on page 54
on page 54
Fault
Pump does not start
Cause
Wiring malfunction.
Remedy
Check and repair wiring.
Cause
Motor protection switch incorrectly set (three-phase motors only).
Remedy
Set motor protection switch properly.
Cause
Motor is malfunctioning.
Remedy
Replace the motor.
Cause
The operating voltage does not match that of the motor.
Remedy
Check whether the correct product number was booked. Adjust the voltage or replace
the motor.
Cause
Oil temperature is below 10 °C (50 °F)
Remedy
Check the temperature with the team. Heat the pump and pump oil or use a different oil.
Cause
Oil is too viscous.
Remedy
Change the oil. Refer to
on page 44.
Cause
Pump is seized.
Remedy
Repair the pump. Contact Service.
Fault
Pump does not reach the ultimate pressure
Cause
Measuring technique or gauge is unsuitable.
Remedy
Use correct measuring technique and gauge. Measure the pressure directly at the intake
port of the pump.
Cause
External leak
Remedy
Repair the pump. Contact Service.