15-11
Options: 1b (calculate once every respiration cycle), 10s
(calculate once every 10 seconds) and 20s (calculate once
every 20 seconds).
[Balance Gas]
: set the balance gas in patient's respiration air
flow.
There are three kinds of selecta
ble balance gas: ―Air‖, ―N2O‖
and ―He‖, namely: air, nitrous oxide and helium. If no specific
balance gas is given, the balance gas can be set as ―Air‖.
[Gas Compensation]
: adjust the concentration of the
compensating gas in patient's respiration air flow. Generally, the
compensating gas is Oxygen, so it can be called oxygen
compensation concentration. Setting range: 1~100; Unit: %;
Default value: 16.
[AG]
: set whether adding the anesthetic gas to patient's
respiration air flow and the concentration of anaesthetic gas.
Setting range: 0.0~20.0; Unit: %; Default value: 0.
The default status is not adding anaesthetic gas, that's to say,
the concentration is 0.0%.
[Atmospheric Pressure]
: set ambient atmospheric pressure.
Setting range: 400~850; Unit: mmHg.
It can be determined by barometer or the ambient altitude.
Altitude can be used to determine the typical barometric
pressure if a barometer is not available, refer to
Typical Pressures and CO2 Readings at Altitudes
15.6.2 Setting CO
2
Alarm Properties
To set CO
2
alarm options, follow the steps below:
1. Enter CO
2
alarm setup screen in either of the following ways:
Select the CO
2
parameter area
→
press the [Alarms]
button.
Press the [Alarms] quick key
→
select the [CO
2
] tab.
Press [Main Menu]
→
[Alarms], select the [CO
2
] tab.
2. Set alarm properties as desired.
15.7 Measurement Limitations
The following factors may influence the measurement accuracy:
Leaks or internal venting of sampled gas.
Mechanical shock.
Cyclic pressure up to 10kPa (100cmH
2
O).
Содержание AIView 12V
Страница 1: ...AIView Series Patient Monitor Operator s Manual...
Страница 2: ......
Страница 6: ...IV This page intentionally left blank...
Страница 24: ...1 7 This page intentionally left blank...
Страница 26: ...2 2...
Страница 31: ...2 7 2 6 2 Rear View 1 2 3 4 1 Handle 2 Device label 3 Equipotential grounding terminal 4 AC power connector...
Страница 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Страница 35: ...2 11 This page intentionally left blank...
Страница 44: ...3 9 1 2 3 4 5 8 9 7 6...
Страница 55: ...3 20 This page intentionally left blank...
Страница 71: ...5 11 This page intentionally left blank...
Страница 78: ...6 7 This page intentionally left blank...
Страница 106: ...8 5 This page intentionally left blank...
Страница 112: ...9 6 This page intentionally left blank...
Страница 119: ...10 7 This page intentionally left blank...
Страница 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Страница 165: ...15 12 Other sources of interference if any...
Страница 166: ...15 13 This page intentionally left blank...
Страница 181: ...17 8 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 185: ...17 12 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 194: ...17 21 This page intentionally left blank...
Страница 204: ...19 5 This page intentionally left blank...
Страница 210: ...20 6 This page intentionally left blank...
Страница 229: ...23 10 This page intentionally left blank...
Страница 248: ...25 8 This page intentionally left blank...
Страница 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Страница 283: ...D 2 This page intentionally left blank...
Страница 289: ...E 6 less than 3V m...